7 Cleaning and sterilization
7
Cleaning and sterilization
After finishing treatment, switch off the laser and disconnect the power cord
from the power supply.
WARNING
The handpieces and the tips are supplied non-sterile, you must sterilize these
products prior to first use and prior to each subsequent use.
WARNING
Disinfect the finger switch and the fiber prior to the first use and prior to each
subsequent use.
Laser protective goggles
Before cleaning the laser protective goggles, please read and observe the
instructions for use provided by the manufacturer and attached to the goggles
in the case.
7.1
Cleaning
NOTICE
i
Manual cleaning must always be combined with disinfection.
NOTICE
i
All tissue residues must be removed from the optical fiber before it is detached
from the handpiece. This prevents internal contamination of the tip.
Following treatment:
Optical fiber
CAUTION
Reattach the black protective cap to the (laser) end of the connector before
taking any hygienic measures.
¾
Remove the optical fiber from the handpiece.
¾
Cut off approx. 4 cm from the distal end of the optical fiber. The cut must
be made perpendicular to the axis of the optical fiber.
¾
Remove approx. 4 cm of the sheath if necessary (see Chapter 4.7.4,
page 27).
¾
The take-up reel and optical fiber must be disinfected first.
¾
After it is removed from the reel, the optical fiber can be cleaned with a
suitable brush under running water (of at least drinking water quality).