8 Maintenance and service
båÖäáëÜ
båÖäáëÜ
8.1.3
Error message and incorrect parameter values
If the following screen appears during one of the three calibration tests:
Repeat the test from the beginning. If the error persists, please contact your
local dental distributor or your authorized service center. Do not, under any
circumstances, continue to use the laser unit!
If the values measured by the power meter are outside the permissible range
in each of the three calibration tests, please proceed according to the follow-
ing flow chart:
Start
One of the three power
values is outside the
permissible range.
Has the optical fiber
been properly
prepared?
Trim the end of the fiber
so that it projects a
well-defined circle.
Does the optical fiber
point exactly to the
sensor area?
Make the optical fiber point
to the measuring head
from a min. distance of
2 cm and ensure that the
red circle is within the
sensor area.
The power
meter is on and set
to correct measuring range
of up to 10 W at 970 nm
± 15 nm.
Set power meter to
measuring range of up to
10 W (upper end of scale)
and wavelength of
970 nm +/– 15 nm.
Does the error
persist when repeating
the measurement?
Sirona Dental
Systems LLC,
Dental distributor
or authorized
customer service
NO
YES
NO
YES
NO
YES
NO
OK