7 Cleaning and sterilization
Approved for sterilization are steam sterilizers that fulfill the requirements of
EN 13060 or at least use fractionated vacuum and are suitable for the steril-
ization of dental handpieces such as e.g. SIRONA DAC PROFESSIONAL.
7.4
Cleaning of the laser unit and finger switch
¾
Use a dry, soft cloth to remove dust from the SIROLaser. More stubborn
spots can be removed with a damp cloth.
For disinfection, you can treat the SIROLaser with all products commonly
used for medical electrical equipment, such as MinutenSpray classic,
CAVICIDE
®
or Caviwipe.
Observe the instructions provided by the manufacturers of these disinfec-
tants.
¾
Disinfect the finger switch by spray or wipe disinfection
You can use for instance: MinutenSpray classic from Alpro; MinutenWipes
from Alpro. In the USA : cavicide
®
and caviwipes
TM
.
Observe the instructions provided by the manufacturers of these disinfec-
tants.
CAUTION
Take care that the fiber will not be damaged in case of disconnecting the finger
switch.
WARNING
For further precautions, see chapter 1.3.
CAUTION
The SIROLaser unit cannot be sterilized.
CAUTION
The handpieces and tips must be
sterilized after each treatment
!