SSG5000A RF Signal Generator User Manua
l
SSG5000A User Manual 82
3.4 Output a pulse train
A pulse train is a user-defined sequence of high and low RF levels with specific time
durations for each part. In this example, we will show how to configure the RF source to
output the following pulse train:
Serial
number
Positive pulse
width
Negative pulse
width
Repetitions
1
10ms
30ms
1
2
20ms
20ms
2
1. Restore factory settings
a) Press
UTILITY
or touch
Preset Type
->
Default
.
b) Press
PRESET
or reset the system settings to perform the reset operation.
2. Install the pulse train option
a) Activate the pulse train generator option.
3. Edit pulse list
a) Press
MOD
or select ANALOG MOD from the Home page. You can swipe left or right
or press
MOD
repeatedly to reach the Pulse Mode screen. Set Pulse Mode to Train,
and the Pulse Train menu will appear.
b) Click
in Pulse Train menu to enter settings, set the positive width of first pulse to
10ms, the negative width to 30ms, and the count to 1.
c)
Click the “+” on the left to add one line, set the positive pulse width to 20ms, the
negative pulse width to 20ms, and the count to 2.
d) After clicking, you can click
on the left side to see the schematic diagram of the
pulse train as follows.