SSG5000A RF Signal Generator User Manua
l
SSG5000A User Manual 30
Click
to enter the file recall page, select and open existing sweep list files.
Please refer to "
" for details.
7) Save
Click
to enter the file save page and save the sweep list to file. Please refer to
" for details.
8) Parameter edit
Click each parameter in the table to set it through the touch screen keyboard or the
keypad of the front panel.
The parameters of each line represent the setting of each sweep point:
a) Frequency: Indicates the frequency of the sweep point.
b) Level: Indicates the level of the sweep point.
c) Time: Indicates the dwell time of the sweep point.
Direction
Sweep direction state is “Up” by default. The SSG5000A provides two types: “Up” or
“Down”. Click the drop-down box to enable the corresponding sweep direction.
⚫
Up: The signal generator sweeps from the start frequency or start level to the stop
frequency or stop level. The progress bar displayed in the parameter bar is swept
from left to right.
⚫
Down: The signal generator sweeps from the stop frequency or stop level to the start
frequency or start level. The progress bar displayed in the parameter bar is swept
from right to left.
Sweep Mode
Default sweep
mode is “Continuous”. The SSG5000A offers “Continuous” or “Single”
sweep modes. Click the drop-down box to enable the desired mode.
⚫
Continuous: When the trigger condition is met, the signal generator is continuously