SSG5000A RF Signal Generator User Manua
l
61 SSG5000A User Manual
8. Power On Line
Set whether the instrument needs to press the power on button to start after the line power
is turned on. In some situations, you may want the instrument to automatically restart if the
line power is restored.
⚫
Off: You need to manually press the power key on the front panel to start the
instrument when the line power is available.
⚫
On: If line power is available, the instrument starts automatically.
Press
UTILITY
->
Setting
->
Power on Line
to switch the state.
9. 10M Adjustment
Modifying the Reference Oscillator Code will change the reference frequency. You can
externally connect a frequency counter from the 10MHz OUT connector on the rear panel
and adjust the 10MHz reference by modifying the code.
⚫
Press
to save
“*.dac” file to store the Ref Oscillator Code you
currently set.
⚫
Press
to recall the Ref Oscillator Code you have stored.
⚫
Press
to set the Ref Oscillator Code to default value, and the
reference frequency will reset to the original value, too.
Press
UTILITY
->
Setting
->
10M Adjustment
to switch the state.
10. Time Setting
Set the display time of the instrument.
Press
UTILITY
->
Setting
->
Time Setting
, then move the cursor to specified location and
change the time by inputting the value.