SSG5000A RF Signal Generator User Manua
l
47 SSG5000A User Manual
Double Pulse Delay
The double pulse delay indicates the delay from the start of the first pulse to the start of
the second pulse in a single cycle of the double pulse modulation signal. This parameter
can be set by touching screen keyboard or pressing the front panel keypad.
Note:
This setting item will be hidden when the modulation source is “Ext” or the pulse
mode is “Single” or “Train”.
#2 Pulse Width
#2 Pulse Width represents the width of the second pulse of the double pulse modulation
signal. This parameter can be set by touching screen keyboard or pressing the front panel
keypad.
Note:
This setting item will be hidden when the modulat
ion source is “Ext” or the pulse
mode is “Single” or “Train”.
Trigger Out
The default is “off”. Click the slide switch to switch the trigger output status.
⚫
On: Turn on the trigger output function. At this time, the signal source outputs the
trigger signal generated by the internal pulse generator from the
[TRIG IN/OUT]
connector
on the rear panel.
⚫
Off: Turn off the trigger output function.
Note:
When the trigger mode is “Ext Trig” or “Ext Gate”, the trigger output function will be
automatically turned off. Only the trigger signal of the PULSE modulation can be output.
The trigger signals of LF SWEEP and RF SWEEP cannot be output.