SSG5000A RF Signal Generator User Manua
l
SSG5000A User Manual 74
refreshes the power sensor measurement value in real time.
⚫
Off: Turn off the power sensor level control function.
Measurement
Indicate the current reading of the sensor. Select the unit used for result display. You can
change the unit for the results disp
lay: dBm, dBμV, uV, mV, V, nW, uW, mW, W.
Target Level
Specify the nominal level expected at the input of the sensor. The signal generator adjusts
the output power accordingly, in order to meet the target value at the sensor input, and
thus the power required at the DUT.
The value can be set by the touch screen keyboard or the front panel keypad.
Limit Level
Set an upper limit for the RF output power to protect your DUT from damage due to high
input power. If the input RF signal power exceeds the limit, the set value will not take effect
and the source will pop up a warning message.
Catch Level
The capture range means that if the power sensor reading is within the effective capture
range, it is a valid reading and the RF signal will be adjusted and compensated. If the
power sensor reading exceeds this range, the reading will be ignored.
Effective Capture Range = Target Level +/- Catch Range