SSG5000A RF Signal Generator User Manua
l
SSG5000A User Manual 42
Use an external modulation signal. The external signal is input via the
[EXT MOD INPUT]
connector on the rear panel of the instrument. The modulated signal can be an arbitrary
waveform.
3. Int + Ext
Use both, the internal and external modulation signal; this technique can be used to
perform dual-tone PM.
PM Shape
Select
MOD
->
PM
->
PM Shape
. After setting the source to “Int” or “Int+Ext”, you can
select PM waveform as “Sine” or “Square”.
PM Rate
Select
MOD
->
PM
->
PM Rate
, then you can set the PM rate.
1. The sine wave modulation frequency range is 0.01Hz~100kHz
2. The square wave modulation frequency range is 0.01Hz~20kHz
PM Deviation
The setting range is
0.00001rad~N × 5rad
. For the definition of N, refer to the data sheet.
⚫
When the modulation source is set to “Int“, the set value is actually the phase offset of
the RF output.
⚫
When the modulation source is set to “Ext”, it is related to the amplitude of the
external input modulation signal.
Offset of RF output = amplitude of input signal ×
external modulation sensitivity
.