When new batteries have been installed, power-up procedure will enable resetting battery
capacity and setting up date and time. Battery capacity reset, date and time can also be
corrected via parameters 508 and 100.
When battery plug is connected, meter will display meter version for 10 seconds. Display will
then show "rESEt.bAt" indicating the option to reset internal battery power calculation. To
execute reset, press key within 6 seconds. If key is not pressed, meter will proceed to set date,
set clock, and finally normal operation mode.
If key is pressed within reset battery time, display will indicate "Accept" to ensure that reset
should take place. Reset will take place only if key is pressed again within the next 6 seconds.
If not, normal operation will begin.
Service and maintenance
8.5 Power up with battery reset, date and time set up
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
79
Содержание SITRANS F M MAG 8000
Страница 10: ...Introduction 1 3 Further Information SITRANS MAG 8000 10 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 16: ...Description 3 4 Benefits SITRANS MAG 8000 16 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 56: ...Commissioning 6 6 Data protection SITRANS MAG 8000 56 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 74: ...Operation 7 6 Battery powered operation SITRANS MAG 8000 74 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 90: ...Troubleshooting FAQs 9 3 Flow simulation SITRANS MAG 8000 90 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 164: ...Index SITRANS MAG 8000 164 Operating Instructions 11 2018 A5E03071515 AE ...