All values are theoretical and are calculated on the assumption that:
● All bolts are new and material selection is according to EN 1515-1 table 2.
● Gasket material not exceeding 75 shore A is used between the water meter and mating 

flanges.

● All bolts are galvanized and adequately lubricated.
● Flanges are made of carbon steel.
● Water meter and mating flanges are correctly aligned.

Torques for 7ME6810 and 7ME6820

Nominal size

PN 10

PN 16

PN 40

Class 150

AWWA

Mm

inch

Nm

f/lbs

Nm

f/lbs

Nm

f/lbs

Nm

f/lbs

Nm

f/lbs

25

1"

N/A

N/A

N/A

N/A

10

7

7

5

N/A

N/A

40

1½"

N/A

N/A

N/A

N/A

16

12

9

7

N/A

N/A

50

2"

N/A

N/A

25

18

N/A

N/A

25

18

N/A

N/A

65

2½"

N/A

N/A

25

18

N/A

N/A

25

18

N/A

N/A

80

3"

N/A

N/A

25

18

N/A

N/A

34

25

N/A

N/A

100

4"

N/A

N/A

25

18

N/A

N/A

26

19

N/A

N/A

125

5"

N/A

N/A

29

21

N/A

N/A

42

31

N/A

N/A

150

6"

N/A

N/A

50

37

N/A

N/A

57

42

N/A

N/A

200

8"

50

37

50

37

N/A

N/A

88

65

N/A

N/A

250

10"

50

37

82

61

N/A

N/A

99

73

N/A

N/A

300

12"

57

42

111

82

N/A

N/A

132

97

N/A

N/A

350

14"

60 

44

120

89

N/A

N/A

225

166

N/A

N/A

400

16"

88

65

170

125

N/A

N/A

210

155

N/A

N/A

450

18"

92

68

170

125

N/A

N/A

220

162

N/A

N/A

500

20"

103

76

230

170

N/A

N/A

200

148

N/A

N/A

600

24"

161

119

350

258

N/A

N/A

280

207

N/A

N/A

700

28"

200

148

304

224

N/A

N/A

N/A

N/A

200

148

750

30"

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

240

177

800

32"

274

202

386

285

N/A

N/A

N/A

N/A

260

192

900

36"

288

213

408

301

N/A

N/A

N/A

N/A

240

177

1000

40"

382

282

546

403

N/A

N/A

N/A

N/A

280

207

1050

42"

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

280

207

1100

44"

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

290

214

1200

48"

395

292

731

539

N/A

N/A

N/A

N/A

310

229

Torques for 7ME6880

Nominal 

size

EN drilled 

pattern

PN 7

ANSI dril‐

led pattern

Class150

AS2091 

drilled pat‐

tern

PN 7

AS4087
PN 16

AS2129 Tabel E
PN 14

Mm

inch

Nm

f/lbs Nm f/lbs Nm

f/lbs Nm

f/lbs

Nm

f/lbs

25

1"

N/A N/A N/A N/A N/A

N/A

N/A N/A

6

4

Installing/Mounting

4.1 Sensor installation

SITRANS MAG 8000

Operating Instructions, 11/2018, A5E03071515-AE

23

Содержание SITRANS F M MAG 8000

Страница 1: ...6810 MAG 8000 7ME6820 MAG 8000 CT 11 2018 A5E03071515 AE Introduction 1 Safety notes 2 Description 3 Installing Mounting 4 Connecting 5 Commissioning 6 Operation 7 Service and maintenance 8 Troubleshooting FAQs 9 Technical data 10 Flow Tool A Qualification certificate B Appendix C ...

Страница 2: ...se who based on their training and experience are capable of identifying risks and avoiding potential hazards when working with these products systems Proper use of Siemens products Note the following WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation If products and components from other manufacturers are used these ...

Страница 3: ... 3 3 Design 13 3 4 Benefits 15 4 Installing Mounting 17 4 1 Sensor installation 18 4 1 1 Locating the sensor 18 4 1 2 Orienting the sensor 20 4 1 3 Mounting the sensor 22 4 2 Potential equalization 25 4 3 Grounding 25 4 4 Cathodic protected pipes 27 4 5 Potting and direct burial 28 4 6 Transmitter installation 29 5 Connecting 31 5 1 General safety requirements 32 5 2 Remote version 33 5 3 Power su...

Страница 4: ...operation 69 8 Service and maintenance 75 8 1 Maintenance 75 8 2 MAG 8000 service guidelines 75 8 3 Replacing transmitter or PCB board 76 8 4 Battery replacement 77 8 5 Power up with battery reset date and time set up 79 8 6 Verification 80 8 7 User and verification sealings 81 8 8 Technical support 82 8 9 Return procedures 83 8 10 Battery disposal 84 9 Troubleshooting FAQs 85 9 1 Fault codes 85 9...

Страница 5: ...2 A 6 Output configuration 125 A 7 Default display information and accessible display menus 125 A 8 Internal data handling 126 A 9 Battery configuration 128 B Qualification certificate 129 B 1 Commissioning 129 B 2 Enabling insulation test 130 B 3 Uploading the device data to the pc 131 B 4 Generating the qualification certificate 132 B 5 Result evaluation 134 C Appendix 135 C 1 Unit conversion ta...

Страница 6: ...C 5 Spare parts Accessories 157 C 5 1 Ordering of spare parts 157 C 6 Features 158 Index 163 Table of contents SITRANS MAG 8000 6 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 7: ...ing instructions are read understood and followed by the relevant personnel before installing the device 1 1 Items supplied SITRANS F M MAG 8000 Standard MAG 8000 CT or MAG 8000 Irrigation Calibration certificate Operating Instructions SITRANS F M documentation disk Inspection 1 Check for mechanical damage due to possible improper handling during shipment All claims for damage are to be made promp...

Страница 8: ...05 2014 NMI M 10 approval for MAG 8000 Irrigation 10 2012 Primary operation software changed from Flow Tool to PDM Flow Tool moved to appendix Qualification Certificate functionality added 03 2010 Integration of MAG 8000 for irrigation applications Restructuring of the document 1 3 Further Information Product information on the Internet The Operating Instructions are available on the documentation...

Страница 9: ...cal contact person http www automation siemens com aspa_app contactmenu aspx ci yes regid DEF lang en Introduction 1 3 Further Information SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 9 ...

Страница 10: ...Introduction 1 3 Further Information SITRANS MAG 8000 10 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 11: ...ion of the equipment must comply with national regulations For example EN 60079 14 for the European Community Instrument safety standards The device has been tested at the factory based on the safety requirements In order to maintain this condition over the expected life of the device the requirements described in these Operating Instructions must be observed NOTICE Material compatibility Siemens ...

Страница 12: ...o the making available on the market of pressure equipment Measuring instruments direc tive MID 2014 32 EU Directive of the European Parliament and the Council on the har monisation of the laws of the Member States relating to the mak ing available on the market of measuring instruments The applicable directives can be found in the EU declaration of conformity of the specific device 2 4 Lithium ba...

Страница 13: ...tomer operation and information on site The transmitter drives the magnetic field in the sensor evaluates the flow signal from the sensor and calculates the volume passing through It delivers the required information via the integrated pulse output or communication interfaces as part of a system solution Its intelligent functionality information and diagnostics ensure optimum meter performance and...

Страница 14: ...tandard compact Remote Figure 3 3 MAG 8000 Standard remote Custody transfer Compact and remote versions Figure 3 4 MAG 8000 CT compact version Description 3 3 Design SITRANS MAG 8000 14 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 15: ...sures minimal pressure loss even at the highest flow rates Overall network system pressures can be reduced helping to prevent leakage from burst pipes and excess stress placed on pumping stations Zero maintenance designed without moving parts and has a 10 year battery life Measurement in both directions only one meter required for measuring in both direction Intelligent meter only one meter for le...

Страница 16: ...Description 3 4 Benefits SITRANS MAG 8000 16 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 17: ...re not exceeded General information This chapter describes how to install the water meter in the compact version as well as in the remote version 87L130 10 Figure 4 1 Compact installation Figure 4 2 Remote installation The installation consists of two steps 1 Sensor installation 2 Transmitter installation remote version only SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 17 ...

Страница 18: ...is present 87L166 N S Flow direction Ensure that sensor is mounted in correct flow direction as indicated on label ORZ GLUHFWLRQ If process flow direction is opposite of flow direction indicated on sensor label forward flow rates can be restored via parameter 327 if factor is adjusted to 1 MAG 8000 CT is approved for bidirectional measurement Installing Mounting 4 1 Sensor installation SITRANS MAG...

Страница 19: ...d 0D of straight pipe downstream from the sensor It is however recommended to follow the installation guidelines for the MAG 8000 Standard water meter for optimal measurement performance Sensor must be completely full of liquid Therefore avoid Air in pipe Installation at the highest point in pipe system Installation in vertical pipes with free outlet Installing Mounting 4 1 Sensor installation SIT...

Страница 20: ... at top where air bubbles may occur and in bottom where mud sludge sand etc may deposit and cover the electrode thus impacting the measurement If Empty Pipe Detection is used sensor should be tilted 45 as shown in upper right figure to maximize full pipe detection and provide accurate volume calculations Note Physical installation of battery pack may influence battery capacity Optimal battery capa...

Страница 21: ...ensor Installation in large pipes MAG 8000 The water meter can be installed between two reducers e g DIN 28545 With an 8 reducer the following pressure drop curve applies The curves are applicable to water Installing Mounting 4 1 Sensor installation SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 21 ...

Страница 22: ...r gasket selection Only use flat rubber gaskets Thickness 1 to 6 mm 0 0 to 0 02 ft dependent on gap tolerance Inner diameter must be larger than bore of water meter Material should be compatible with process fluid Hardness should be maximum 75 Shore A Maximum allowable torques Standard bolts must be well lubricated and tightened evenly around gasket Leakage damage to water meter or piping may aris...

Страница 23: ...8 65 N A N A 250 10 50 37 82 61 N A N A 99 73 N A N A 300 12 57 42 111 82 N A N A 132 97 N A N A 350 14 60 44 120 89 N A N A 225 166 N A N A 400 16 88 65 170 125 N A N A 210 155 N A N A 450 18 92 68 170 125 N A N A 220 162 N A N A 500 20 103 76 230 170 N A N A 200 148 N A N A 600 24 161 119 350 258 N A N A 280 207 N A N A 700 28 200 148 304 224 N A N A N A N A 200 148 750 30 N A N A N A N A N A N ...

Страница 24: ...32 24 40 30 51 38 153 113 N A N A 400 16 46 34 50 37 62 46 172 127 N A N A 450 18 47 35 56 42 79 59 224 165 N A N A 500 20 57 43 67 50 72 54 198 146 N A N A 600 24 89 66 104 77 111 82 287 211 N A N A 700 N A N A N A N A N A N A N A 228 168 N A N A 750 N A N A N A N A N A N A N A N A N A N A N A 800 N A N A N A N A N A N A N A 426 314 N A N A 900 N A N A N A N A N A N A N A 416 307 N A N A 1000 N A...

Страница 25: ...ation 7ME6880 4 3 Grounding The sensor body must be grounded using grounding bonding straps and or grounding rings to protect flow signal against stray electrical noise and or lightning This ensures that noise is carried through sensor body and that the measuring area within sensor body is noise free Irrigation applications MAG 8000 Irrigation device is always delivered with pre mounted grounding ...

Страница 26: ... part of delivery and pre mounted on water meter Plastic pipes and lined metal pipes Use optional grounding rings at both ends Plastic pipes or lined metal pipes Grounding rings are not included in delivery Installing Mounting 4 3 Grounding SITRANS MAG 8000 26 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 27: ... heavier copper wire and connected with 6 mm screws 4 4 Cathodic protected pipes Pay special attention to meter installation in cathodic protected pipe Isolate meter from pipeline by mounting isolation sleeves and washers on flange bolts and connect a wire dimensioned to manage the cathodic current and environmental influence between pipelines Installing Mounting 4 4 Cathodic protected pipes SITRA...

Страница 28: ...e 2 Mount O ring properly and correctly and grease with gel 3 Fill Sylgard potting kit in bottom part of casing 4 Renew Silicagel bag placed behind battery cup to prevent condensation within meter if necessary 5 Mount the enclosure carefully and make sure not to damage the O ring Note Potting Ensure not to fill Sylgard potting kit in the space for the battery pack Ensure Silicagel bag is not in co...

Страница 29: ...avation be necessary Before covering pea gravel with earth use electrical cable identification tape above gravel Run remote sensor cable through a plastic conduit of minimum 50 mm 2 inches 4 6 Transmitter installation Mount bracket on a wall as shown below or on a horizontal or a vertical pipe using ordinary hose clips or duct straps Wall mounting Figure 4 3 Wall mounting Installing Mounting 4 6 T...

Страница 30: ...Pipe mounting Figure 4 4 Pipe mounting vertical Figure 4 5 Pipe mounting horizontal Installing Mounting 4 6 Transmitter installation SITRANS MAG 8000 30 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 31: ...tion Output A Output B The 3 6 V DC battery connector male and pulse connection terminals are placed in the right side of the PCB board see figure Connection for add on interface modules is placed on the left side HL Hardware lock key connection V Verification button enables verification mode for 4 hours To configure the outputs via PDM see Output configuration Page 53 To configure the outputs via...

Страница 32: ...nnected to the power supply If the housing is under voltage power supply the cover may be unscrewed by qualified personnel only WARNING Mains supply from building installation Class II A switch or circuit breaker max 15 A must be installed in close proximity to the equipment and within easy reach of the operator It must be marked as the disconnecting device for the equipment Connecting 5 1 General...

Страница 33: ...IUS of 45 mm 1 8 inches otherwise there s a risk of damaging the cable Save dust covers for future use and protection 3 Ensure connectors are clean 4 Ensure connectors are fastened securely to achieve a good connection and watertight seal Note If dirt enters connector ends use plain water for cleaning Ensure connectors are completely dry before making connections For MAG 8000 CT user sealing see U...

Страница 34: ... be connected to male connector 3 6 V DC on PCB board Battery backup input Male battery connector with black and red wires black wire is ground Male battery connector has to be connected to female connector on backup battery Functional ground Black wire with terminal must be connected to the device encapsulation with a screw Mains power supply has to be connected to a switch near water meter accor...

Страница 35: ...ulation with a screw Cable installation Choose the correct glands for the selected cable type see Accessories Page 157 for glands selection Ensure shield is mounted under cable clamps do not make pig tails Note Mains or line powered PUR cable no shield has to be mounted under cable clamps All cable glands have to be sufficiently tightened to ensure IP rating Connecting 5 3 Power supply SITRANS MAG...

Страница 36: ... up see Commissioning Page 53 Pulse output is not polarized and can be connected for positive or negative logic Pull up down resistor R is selected in relation to power supply voltage V and with a max current I of 50 mA Note Pulse output isolation The MAG 8000 pulse output has a functional isolation only and must therefore be connected to equipment complying with the Low Voltage Directive in order...

Страница 37: ...ve ground Common A Modbus over serial line cable must be shielded At one end of each cable its shield must be connected to protective ground If a connector is used at this end connector shell must be connected to cable shield Connecting 5 5 Communication modules SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 37 ...

Страница 38: ...eld must always be connected to MAG 8000 encapsulation using cable clamp as shown in figure in section cable installation see Power supply Page 34 Bus termination All RS 485 based networks must be terminated correctly to function properly A termination must be placed at each end of segment The Modbus RTU module can add a 120 ohm termination if a jumper is placed beside terminals in position ON Ter...

Страница 39: ...o terminal 93 WARNING Unshielded wire It is important that unshielded wire does not touch any metal parts of MAG 8000 housing Other radio interface cable has to be a 3 wire cable with a shield connected to MAG 8000 housing mounting cable shield is shown to the right GND Data 129 10 10 Power clock Connect shield to encapsulation Radio interface 128 10 10 Connecting 5 5 Communication modules SITRANS...

Страница 40: ...nternet at www siemens com flowdocumentation www siemens com flowdocumentation Information on the operation of communication modules For information on how to operate the various communication modules see GSM GPRS Operating Instructions http support automation siemens com WW view en 60545899 Modbus RTU Operating Instructions http support automation siemens com WW view en 54091509 Connecting 5 6 Co...

Страница 41: ...osis signals of device This chapter describes how to set up SIMATIC PDM and how to configure the device for commissioning Note For instructions on installation and operation of SIMATIC PDM refer to the SIMATIC PDM Getting Started 6 2 Initial commissioning via SIMATIC PDM This chapter describes how to install the PDM device driver EDD The newest EDD driver is available on the internet http support ...

Страница 42: ...h where the file was downloaded and saved The PDM will show a tree with the Software options Click on the Siemens checkbox to select all options Click OK to install the new EDD files Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS MAG 8000 42 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 43: ...are the same only the views on the PC screen differ Setting up meter 1 Click on File 2 Select New 3 Assign a new project name in the Name box e g MAG8000 4 In Process Device Network View right click on Net select Insert New Object MODBUS net Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 43 ...

Страница 44: ...5 Right click on MODBUS net select Object Properties Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS MAG 8000 44 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 45: ... 8000 is 19 200 baud If the IrDA adapter is connected properly a small IrDA icon will appear in the Windows notification area When placing mouse at this icon the device information will be shown e g MAG8000 SN1234 is in range Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 45 ...

Страница 46: ...ndow click on Flow Electromagnetic Siemens AG MAG8000 9 Select the proper device MAG 8000 Advanced MAG 8000 Basic MAG 8000 CT Advanced or MAG 8000 CT Basic depending on transmitter version to be controlled Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS MAG 8000 46 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 47: ...e click created device e g MAG8000 Advanced A new window will pop up for device parameter configuration Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 47 ...

Страница 48: ...s Settings to specify whether you are permanently registered as a Maintenance engineer or Specialist and or whether to disable the display of the User dialog box Refer to SIMATIC PDM help manual for more details of Maintenance engineer and Specialist Proceed as follows Click the appropriate button Maintenance engineer or Specialist If the option Specialist is selected enter the password in the Pas...

Страница 49: ... off line data which are not stored in the MAG 8000 yet whereas black texts show the actual meter data Each adjustable parameter has a Default Value a Minimum Value and a Maximum Value which are prompted by right clicking on the parameter More information will be available when the Help button is clicked Commissioning 6 3 Setting the basic parameters SITRANS MAG 8000 Operating Instructions 11 2018...

Страница 50: ...e marked alarms enabled Click on Transfer to instantly synchronize alarm configurations between the PC and the device Figure 6 1 Alarm status marked alarms are enabled Select Device Download to Device to download all the changed parameters to the device Commissioning 6 3 Setting the basic parameters SITRANS MAG 8000 50 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 51: ... right click on the PDM object to be exported select Export In the subsequent dialog box specify the type of export as well as the name and location where the PDM file containing the exported data is to be saved Confirm the entries by clicking OK Commissioning 6 3 Setting the basic parameters SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 51 ...

Страница 52: ...ed showing export progress After the export is completed the status Result is displayed See also www siemens com flow www siemens com flow Commissioning 6 3 Setting the basic parameters SITRANS MAG 8000 52 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 53: ...t be stored into the transmitter via Download to Device 6 5 Output configuration Pulse output can be configured as volume pulse alarm or call up Default factory setting is with output A enabled for forward flow and output B for alarm output Output configuration in SIMATIC PDM Select Output The software window shows the parameters for output configuration Each parameter has its own guideline for se...

Страница 54: ...See Output characteristics Page 96 for more information on how the outputs work Commissioning 6 5 Output configuration SITRANS MAG 8000 54 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 55: ...ctor 302 Pipe size 321 Calibration date 323 Calibration factor 325 Sensor offset 332 Max sensor excitation frequency Additional protected parameters MAG 8000 CT Parameter no Parameter name 101 Totalizer 1 102 Totalizer 2 303 Operation excitation frequency 305 Decimal point place 310 Flow direction totalizer 1 312 Flow direction totalizer 2 327 Adjustment factor 328 Low flow cut off 332 Empty pipe ...

Страница 56: ...Commissioning 6 6 Data protection SITRANS MAG 8000 56 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 57: ...rmation see menu overview Page 62 If key is not pressed for 10 minutes display will time out and return to default configured operator menu Key There are three different ways the interface key will respond to being pressed 1 A brief press less than 2 seconds will advance screen to next index or menu 2 A short press 2 to 5 seconds will enter a menu or escape menu selection 3 A long press more than ...

Страница 58: ...uch key s self calibration once the MAG 8000 transmitter enclosure is removed and mounted again There will be no response from the touch key during self calibration 7 2 Display symbols Top area of display shows status bar 6WDWXV EDU FWXDO WDULII DFFRXQW DQG UHVHW LQGLFDWLRQ XUUHQW DFWLYH PSW SLSH DFWLYH 0DLQV SRZHU VXSSO DWWHU SRZHU VXSSO DQG HQHUJ VWDWXV Figure 7 3 Status bar Status information s...

Страница 59: ...XH LQGLFDWLRQ HIDXOW RSHUDWRU PHQX 0HWHU LQIRUPDWLRQ PHQX 6HUYLFH PHQX DWD ORJJHU PHQX 6WDWLVWLFV PHQX DGYDQFHG YHUVLRQ RQO 7DULII PHQX DGYDQFHG YHUVLRQ RQO 0HQX EDU Figure 7 4 Menu bar The menu bar icons indicate actual selected menu and related index for selected information Display overview shows relation between menu index and information Only operator menu has information and functions that c...

Страница 60: ...es accessible display menus with selection of one or more of the following menus Operator menu Meter info menu Service menu Data logger menu Statistic menu advanced version only Revenue menu advanced version only Operation 7 3 Default display information and accessible display menus SITRANS MAG 8000 60 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 61: ...Operation 7 3 Default display information and accessible display menus SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 61 ...

Страница 62: ...7 4 Operator menu The operator menu consists of several indexes described in the following Operation 7 4 Operator menu SITRANS MAG 8000 62 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 63: ... 3DULW 6WRS ELW Figure 7 5 Menu overview Operation 7 4 Operator menu SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 63 ...

Страница 64: ... menu Totalizer 2 Flow volume totalizer 2 factory configured for reverse flow A negative value indicates reverse flow calculation For MAG 8000 the value of totalizer 2 can be reset to zero or set to any value desired via PDM or Flow Tool example replacing an existing old meter Index 3 Flow rate Figure 7 8 Operator menu Flow rate Index 3 shows actual flow rate If a negative value is indicated flow ...

Страница 65: ...ion and service guidelines are made in service section After all faults have disappeared display shows total hours of faults until meter was reset Fault information Each number indicates a dedicated fault 1 Insulation fault 2 Coil current fault 3 Amplifier overload fault 4 Data base checksum fault 5 Low power warning alarm limits are configurable 6 Flow overload Qmax Q4 125 Qn flow overload 7 Puls...

Страница 66: ...he totalizer is neither configurable nor resttable Display test Figure 7 11 Operator menu Display test All segments of display are alternately flashed on and off during this test Menu selection 6KRUW SUHVV IRU HQWHU HVFDSH PHQX DLW IRU PHQX V PERO WR IODVK VHF Figure 7 12 Operator menu Menu selection If key is pressed shortly 2 to 5 seconds menu selection will flash indicating that a new selection...

Страница 67: ... IRU V PERO WR IODVK VHF 5HVHW DFFHSWHG 5HVHW SRVVLEOH Figure 7 13 Operator menu Call up reset Call up reset window index 0 is only shown when call up function is activated r indicates that it can be reset by a long press on key When releasing key while r is flashing call up function will be reset and window disappears Operation 7 4 Operator menu SITRANS MAG 8000 Operating Instructions 11 2018 A5E...

Страница 68: ...l data handling Meter status Meter status parameter 120 gives a fast indication of reliability of revenue data Operation 7 5 Internal data handling SITRANS MAG 8000 68 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 69: ... made and so on The consumption alarm indicates that the actual consumption on totalizer 1 is above or below the consumption limits Note Invalid log date If the log date 00 00 2000 is shown in the device display the log period was never set 7 6 Battery powered operation The MAG 8000 Standard 7ME6810 and MAG 8000 CT 7ME6820 are factory configured for 6 years and MAG 8000 Irrigation 7ME6880 for 4 ye...

Страница 70: ...suming elements and the influence of temperature changes Consumption and operation time calculation Battery operation time depends on connected battery pack as well as operation conditions of meter Every 4 hours the advanced power management system calculates the real power consumption and remaining operation capacity Power consumption calculation includes flow measurement meter dialog communicati...

Страница 71: ... 6 years 40 months 8 months 4 months 2 months DN 200 to 600 8 to 24 8 years 6 years 4 years 20 months 4 months 2 months N A DN 700 to 1200 28 to 48 6 years 4 years 2 years 1 year 2 months N A N A Four D Cell battery 66 Ah External bat tery pack DN 25 to 150 1 to 6 N A 10 years 10 years 80 months 16 months 8 months 4 months DN 200 to 600 8 to 24 N A 10 years 10 years 40 months 8 months 4 months N A...

Страница 72: ...ery DN 25 to 600 1 to 24 52 months 40 months 25 months 12 months 2 months 1 month DN 700 to 1200 28 to 48 36 months 24 months 12 months 6 months 1 month N A Two D Cell bat tery 33 Ah Internal battery pack DN 25 to 600 1 to 24 105 months 80 months 50 months 24 months 4 months 2 months DN 700 to 1200 28 to 48 72 months 48 months 24 months 12 months 2 months N A Four D Cell bat tery 66 Ah External ba...

Страница 73: ...are located in Alarm group e g Battery limit parameter 206 which is the level at which low power alarm is activated generating an alarm or call up if configured as well as low power fault related parameters 230 to 234 When switching between battery power from the internal and the external battery packs Battery power parameter 507 must be adjusted to match the actual number of batteries connected O...

Страница 74: ...Operation 7 6 Battery powered operation SITRANS MAG 8000 74 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 75: ...tion proper installation and proper commissioning for the particular application This service guideline section indicates how to detect and solve the most common problems Meter and application problems are indicated by the alarm program via the main fault and warning symbol on the display and the comprehensive data logging and monitoring available via the communication interface Alarm monitoring i...

Страница 76: ...ct label Final configuration is done on site Missing data and configuration can be uploaded from the old meter or it can be read from the old meter product label 3 Order only a replacement PCB board The PCB board can only be ordered for basic or advanced version with default settings When making the configuration on site the service mode must be selected in the Flow Tool SIMATIC PDM and the hardwa...

Страница 77: ...ilica gel bag 4 Replace O ring to ensure continued IP68 enclosure rating Check O ring for damage or deformity Smear O ring with acid free lubricating gel 5 Push locking tab and loosen strip 2OG 1HZ 1HZ 6 Remove battery pack with power still connected Service and maintenance 8 4 Battery replacement SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 77 ...

Страница 78: ...tact with the Sylgard 9 Disconnect old battery pack and connect new one 10 Press key within 6 seconds to reset battery i e operating time and remaining capacity when display shows 11 Mount top lid 12 Fasten screws to reassemble meter completely If necessary adjust time and date via software see Power up with battery reset date and time set up Note All sealed MAG 8000 CT meters have to be re verifi...

Страница 79: ... for 10 seconds Display will then show rESEt bAt indicating the option to reset internal battery power calculation To execute reset press key within 6 seconds If key is not pressed meter will proceed to set date set clock and finally normal operation mode If key is pressed within reset battery time display will indicate Accept to ensure that reset should take place Reset will take place only if ke...

Страница 80: ...idth is set to 0 5 ms When verification mode is exited the previous pulse setting is restored Pulse widths other than 1 ms can be selected by storing new pulse values This setting remains when verification mode is exited Activation of verification mode Verification mode is enabled in one of the following ways Pressing verification button through hole in front screen or Writing integer 1 to paramet...

Страница 81: ...ly stops if not manually deactivated within 4 hours 8 7 User and verification sealings MAG 8000 CT must be sealed to remain custody transfer approved Adequate sealing of sensor electronics Compact version Figure 8 2 MAG 8000 CT user sealings A Figure 8 3 MAG 8000 CT verification sealing A Service and maintenance 8 7 User and verification sealings SITRANS MAG 8000 Operating Instructions 11 2018 A5E...

Страница 82: ...e Support on the Internet In addition to our documentation we offer a comprehensive knowledge base online on the Internet at AUTOHOTSPOT There you will find The latest product information FAQs downloads tips and tricks Our newsletter providing you with the latest information about your products Our bulletin board where users and specialists share their knowledge worldwide You can find your local c...

Страница 83: ...e parts returned reason for the return Declaration of Decontamination Declaration of Decontamination http pia khe siemens com efiles feldg files Service declaration_of_decontamination_en pdf With this declaration you certify that the returned products spare parts have been carefully cleaned and are free from any residues If the device has been operated together with toxic caustic flammable or wate...

Страница 84: ...posal services Waste industrial batteries from our products are accepted back by Siemens and by the local Siemens representatives Please follow the return procedures of Siemens or talk to your local Siemens partner Service and maintenance 8 10 Battery disposal SITRANS MAG 8000 84 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 85: ...ulation test parameter 800 will reset alarm and start a new insu lation test During the 4 minute insulation test meas urement is disabled Alarm remains active until a new test period detects no failures 2 Coil current fault Error in the coil circuit Coil current has not set tled within specified pe riod A short circuit in the coil can NOT be detected Coils have been disconnected could be caused by...

Страница 86: ... of Q3 Wrong sizing flow sensor too small Check meter sizing for actual installation Alarm remains active until fault condition is resolved 7 PulseA over load Duty cycle of output A has exceeded maxi mum possible of 50 Wrong settings for output A Change volume per pulse to a higher value see Technical data Page 96 for pulse selection Reminder basic version is limited to 50 Hz maximum advanced vers...

Страница 87: ...Flow Limit Reverse flow rate is be low a preset threshold default 1E9 Water network failure non return valve is broken Alarm remains active until reverse flow rate drops below reverse flow alarm limit Note Reset of fault log parameter 204 also resets all alarms Once reset only active alarms become visible again 9 2 Built in functions Empty pipe detection Electrode impedance is measured with 800 Hz...

Страница 88: ...st When H bridge is turned and just before making samples 4 times each measurement the coil current is checked via a comparator If the coil current is not settled a failure is reported During coil current failure the flow value is forced zero Amplifier test Overload is possible both from the AD conversion of sensor signal and from the pre amplifier These checks are made at each sample and if one o...

Страница 89: ...lly including all other diagnosis functions in the MAG 8000 and the MAG 8000 CT 9 3 Flow simulation MAG 8000 and MAG 8000 CT has a built in flow simulator parameters 551 and 552 to verify and adjust pulse output to any connected device or system WARNING Totalized values are changed during simulation and actual flow is NOT measured Simulation will stop automatically in 4 hours For MAG 8000 CT flow ...

Страница 90: ...Troubleshooting FAQs 9 3 Flow simulation SITRANS MAG 8000 90 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 91: ... 158 F 0 1 to 50 C 32 to 122 F Storage 40 to 70 C 22 to 158 F 40 to 70 C 22 to 158 F 40 to 70 C 22 to 158 F Enclosure IP68 NEMA 6P rating Cable glands mounted requires Sylgard potting kit to remain IP68 NEMA 6P otherwise IP67 NEMA 4 rating is obtained Factory mounted cable provides IP68 NEMA 6P rating Approvals Drinking water ap provals NSF 61 cold water USA WRAS BS 6920 cold wa ter UK ACS Listed ...

Страница 92: ...e D Drilled flanges maximum pressure 7 bar AS 2191 Table E DN25 DN40 DN125 Max excitation fre quency Basic version Battery powered 1 15 Hz for sensor size DN 25 to 150 1 to 6 1 30 Hz for sensor size DN 200 to 600 8 to 24 1 60 Hz for sensor size DN 700 to 1200 28 to 48 1 15 Hz for sensor size DN 25 to 150 1 to 6 1 30 Hz for sensor size DN 200 to 600 8 to 24 1 60 Hz for sensor size DN 700 to 1200 28...

Страница 93: ...ansmitter Transmitter MAG 8000Standard 7ME6810 MAG 8000 Irrigation 7ME6880 MAG 8000 CT 7ME6820 Installation Integral compact or remote with factory mounted cable in 2 5 10 20 or 30 m lengths with IP68 NEMA 6P connectors Connection is made at the transmitter bottom Material Top housing Stainless steel AISI 316 Bottom Coated brass Wall mounting bracket Stainless steel AISI 304 Display and key Displa...

Страница 94: ...50 mA short circuit protected Output A Programmable as Pulse volume Forward Reverse Forward net Reverse net Output B Programmable as Pulse volume Forward Reverse Forward net Reverse net Alarm Call up Pulse rate Basic version Max 50 Hz Advanced version Max 100 Hz Pulse width 5 10 50 100 500 ms Communica tion IrDA Standard integrated infrared communication interface with MODBUS RTU protocol Add on m...

Страница 95: ...o 230 V AC power supply Factory mounted PUR cable 2 x 1 mm2 brown blue Length 3 m Resistance Sunlight and water Outer diameter 7 mm 0 28 Rated voltage 300 500 V AC Testing voltage 2000 V AC Temperature range Fixed laying Flexible application 40 to 90 C 40 to 194 F 30 to 80 C 22 to 176 F Bending radius Min 28 mm fixed installation Pulling force Max 200 N Output Female connector Backup battery Male ...

Страница 96: ...rd Modbus over serial line v1 0 Certified No Device profile None Add on modules comply with MODBUS over serial line specification implementation guide v 1 0 modbus org 12 02 02 MODBUS application protocol specification v 1 1 modbus org 12 06 02 Isolation 500 V AC functional isolation of data signals and common Add on modules must be connected to equipment complying with Low Voltage Directive in or...

Страница 97: ...should not exceed the pulse rate of the output configuration table MAG 8000 CT The volume per pulse is defined in the MAG 8000 CT version to 0 001 or 0 01 or 0 1 or 1 m3 per pulse Output B as alarm output Alarm output Alarm status On Off No Error MAG 8000 When output B is configured as an alarm output it will follow the internal alarms that were previously chosen in the Alarm Configuration List Pa...

Страница 98: ...d MAG 8000 When output B is configured as an call up output it will follow the internal alarms that were previously chosen in the Alarm Configuration List Note Call up output is inverted to a pulse output providing an alarm if power disappears or cable connection is interrupted Factory regional settings DN mm size inch Pulse width ms Europe m3 USA Gallons Australia ML 25 40 50 1 1 2 50 0 01 1 0 00...

Страница 99: ...0 0035 0 018 4 623 0 000018 0 035 0 175 200 8 1000 0 0028 0 0056 0 028 7 338 0 000028 0 056 0 278 250 10 1600 0 0044 0 0089 0 044 11 741 0 000044 0 089 0 444 300 12 2500 0 0069 0 0139 0 069 18 345 0 000069 0 139 0 694 350 14 3463 0 0096 0 0192 0 096 25 412 0 000096 0 192 0 962 400 16 4523 0 0126 0 0251 0 126 33 190 0 000126 0 251 1 256 450 18 5725 0 0159 0 0318 0 159 42 010 0 000159 0 318 1 590 50...

Страница 100: ... over time the net pulse output indicates the bi directional totalizer as calculated internally The same principle applies for forward and reverse flow calculation By changing the status of the pulse output the internal pulse calculator will be reset Flow Net totalizer in meter display Bi directional Pulse output forward Uni directional mode Volume m3 Pulse output net forward Bi direction al mode ...

Страница 101: ... 0 0 0 0 0 0 0 0 630 1 7ME 6820 4P DN200 8 m3 0 0 0 0 0 0 0 0 1000 1 7ME 6820 4V DN250 10 m3 0 0 0 0 0 0 0 0 1600 1 7ME 6820 5D DN300 12 m3 0 0 0 0 0 0 0 0 1600 1 Parameter 305 Decimal Point should be configured to 1 digit after decimal point in case Encoder interface module is selected in MAG 8000 MAG 8000 CT and MAG 8000 Irrigation otherwise proper reading from the device cannot be performed 10 ...

Страница 102: ...ded calibration of the MAG 8000 7ME6810 sizes DN50 2 to DN 300 12 results in a max 0 2 of rate 2 mm s uncertainty A calibration certificate is included with every sensor and calibration data is stored within the meter ORZUDWH P V UURU s MAG 8000 Irriga tion Standard calibra tion 0 8 of rate 2 5 mm s MAG 8000 Stand ard Standard calibra tion 0 4 of rate 2 mm s MAG 8000 Stand ard Extended calibra tio...

Страница 103: ...lass II for the sensor program from DN 50 to DN 300 at different Q3 and Q3 Q1 Q2 Q1 1 6 and follows standard OIML R49 specification O O 2 0 5 ODVV ODVV OIML R49 specification for Class I Size 50 2 65 2 80 3 100 4 125 5 150 6 200 8 250 10 300 12 R Q3 Q1 250 250 250 250 250 250 250 250 125 Q4 m3 h 78 75 125 200 312 5 500 787 5 1250 2000 2000 Q3 m3 h 63 100 160 250 400 630 1000 1600 1600 Q2 m3 h 0 4 ...

Страница 104: ...roduction MAG 8000 CT MI 001 verified and labelled products at a given Q3 and Q4 Q3 1 25 and Q2 Q1 1 6 measuring ranges se below table Size 50 2 65 2 80 3 100 4 125 5 150 6 200 8 250 10 300 12 350 14 400 16 450 18 500 20 600 24 R Q3 Q1 25 25 25 25 25 25 25 25 25 25 25 25 25 25 Q4 m3 h 20 31 25 50 78 75 125 200 312 5 500 787 5 787 5 1250 1250 2000 2000 Q3 m3 h 16 25 40 63 100 160 250 400 630 630 10...

Страница 105: ...00 1600 6300 6300 Q2 m3 h 0 4 0 63 1 1 6 2 5 4 6 3 10 16 16 16 63 63 Q1 m3 h 0 25 0 39 0 63 1 1 56 2 5 3 94 6 25 10 10 10 39 39 Size 50 2 65 2 80 3 100 4 125 5 150 6 200 8 250 10 300 12 350 14 400 16 450 18 500 20 600 24 R Q3 Q1 200 200 200 200 200 200 200 200 200 Q4 m3 h 50 78 75 125 200 312 5 500 787 5 1250 2000 Q3 m3 h 40 63 100 160 250 400 630 1000 1600 Q2 m3 h 0 32 0 5 0 8 1 28 2 3 2 5 04 8 1...

Страница 106: ...0 0 10 0 10 0 8 2 PN 16 10 0 16 0 16 0 13 2 ANSI 16 5 150 lb 10 0 19 7 19 7 16 2 Imperial Pressures in Psi Flange spec Flange rating Temperature F 32 50 122 158 Sizes 1 1 and 12 EN 1092 1 PN 10 145 145 141 136 PN 16 232 232 225 219 PN 40 580 580 561 547 ANSI 16 5 150 lb 286 286 280 261 Sizes 2 to12 EN 1092 1 PN 10 145 145 145 119 PN 16 145 232 232 191 ANSI 16 5 150 lb 145 286 286 235 Technical dat...

Страница 107: ... 98 2 6 13 40 1 204 8 1 20 0 7 9 200 200 N A N A 40 1 57 2 9 20 50 2 195 7 7 200 7 9 200 N A N A 42 1 65 2 11 25 65 2 201 8 200 7 9 200 N A N A 55 2 17 2 13 29 80 3 207 8 2 200 7 9 200 N A N A 67 2 64 2 15 34 100 4 214 8 5 250 9 8 250 N A N A 81 3 19 2 17 38 125 5 224 8 9 250 9 8 250 250 N A N A 101 3 98 2 22 50 150 6 239 9 5 300 11 8 300 N A N A 131 5 16 2 28 63 200 8 264 10 5 350 350 13 8 350 N ...

Страница 108: ...700 27 55 2 31 6 694 750 30 571 22 5 N A N A N A N A 750 29 5 750 29 52 2 N A N A 800 32 606 23 9 800 1000 N A 800 800 31 5 800 31 49 2 39 8 1045 900 36 653 25 7 900 1125 N A 900 900 35 4 900 35 42 2 47 6 1045 1000 40 704 27 7 1000 1250 N A 1000 1000 39 4 1000 39 36 2 60 2 1322 1050 42 704 27 7 N A N A N A N A 1050 41 3 1050 41 33 2 N A N A 1100 44 755 29 7 N A N A N A N A 1100 43 3 1100 43 30 2 N...

Страница 109: ... 500 670 620 38 26 20 M24 600 780 725 42 30 20 M27 700 895 840 30 24 M27 750 800 1015 950 33 24 M30 900 1115 1050 33 28 M30 1000 1230 1160 36 28 M33 1050 1100 1200 1455 1380 39 32 M36 PN 16 50 165 125 19 18 4 M16 65 185 145 20 18 8 M16 80 200 160 20 18 8 M16 100 220 180 22 18 8 M16 125 250 210 22 18 8 M16 150 285 240 24 22 8 M20 200 340 295 26 22 12 M20 250 405 355 29 26 12 M24 300 460 410 32 26 1...

Страница 110: ... Bolting Size D PCD T B Holes Bolts ANSI Class 150 1 4 25 3 12 0 56 0 62 4 9 16 1 5 3 88 0 68 0 62 4 9 16 2 6 4 75 0 75 0 75 4 5 8 2 7 5 5 0 88 0 75 4 5 8 3 7 5 6 0 94 0 75 4 5 8 4 9 7 5 0 94 0 75 8 5 8 5 10 8 5 0 94 0 88 8 3 4 6 11 9 5 1 0 88 8 3 4 8 13 5 11 75 1 12 0 88 8 3 4 10 16 14 25 1 19 1 00 12 7 8 12 19 17 1 25 1 00 12 7 8 14 21 18 75 1 38 1 12 12 1 16 23 5 21 25 1 44 1 12 16 1 18 25 22 7...

Страница 111: ...D T B Holes Bolts ANSI Class 150 drilling pattern 2 6 4 75 0 32 0 75 4 5 8 2 7 5 5 0 32 0 75 4 5 8 3 7 5 6 0 40 0 75 4 5 8 4 9 7 5 0 40 0 75 8 5 8 5 10 8 5 0 52 0 88 8 3 4 6 11 9 5 0 52 0 88 8 3 4 8 13 5 11 75 0 52 0 88 8 3 4 10 16 14 25 0 63 1 00 12 7 8 12 19 17 0 76 1 00 12 7 8 14 21 18 75 1 38 1 12 12 1 16 23 5 21 5 1 44 1 12 16 1 18 25 22 75 1 56 1 25 16 1 1 8 20 27 5 25 1 69 1 25 20 1 1 8 24 ...

Страница 112: ...20 DN40 135 98 12 7 14 4 M20 DN125 255 210 14 18 8 M20 Dimensions mm Bolting Size D PCD T B Holes Bolts AS4087 PN16 DN50 150 114 11 61 5 4 M16 DN65 165 127 11 77 5 4 M16 DN80 185 146 11 90 5 4 M16 DN100 215 178 13 116 4 M16 DN150 280 235 13 170 5 8 M16 DN200 335 292 19 221 5 8 M16 DN250 405 356 19 275 5 8 M20 DN300 455 405 23 326 5 12 M20 DN350 525 470 30 358 12 M24 DN600 825 756 48 613 16 M27 DN8...

Страница 113: ...sition as shown The cable for the battery pack connection must be ordered separately Worn industry batteries may be disposed at the manufacturer or the importer who originally marketed the battery or where new batteries are bought Grounding rings Sizes DN 25 to 300 7ME6810 and 7ME6820 Figure 10 2 Flat ring Technical data 10 12 Dimensions and drawings SITRANS MAG 8000 Operating Instructions 11 2018...

Страница 114: ...grounding ring factory pre mounted Sizes DN 350 to 600 7ME6810 and 7ME6820 and DN 700 to 1200 7ME6810 W W Figure 10 4 Type C grounding ring Technical data 10 12 Dimensions and drawings SITRANS MAG 8000 114 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 115: ...pport homepage and the latest downloads Device driver The Device Drivers are related to meter version and is automatically selected in Automatic mode In Manual mode meter version is manually selected and version check is automatically made when data is uploaded or downloaded Data backup Meter information is stored in an internal data prom that secures data in case of power failure Information like...

Страница 116: ... Double click on Flow Tool icon on windows desktop to start program Note For on line communication make sure the IrDA symbol is visible in the toolbar status menu before starting Flow Tool A 3 Configuring the device This chapter describes how to set up meter for communication with PC The options in both Flow Tool and SIMATIC PDM software are the same only the views on the PC screen differ Flow Too...

Страница 117: ...Right click on Project 2 Select New 3 Select Flow Meter 4 Name device 5 Select configuration type Automatic or Manual Flow Tool A 3 Configuring the device SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 117 ...

Страница 118: ...wnloaded to the meter after wards A 4 Setting the basic parameters Note No factory reset The device is delivered with factory settings that are not stored as default values Because no default values are present in the meter an automatic return to factory values is not possible Flow Tool A 4 Setting the basic parameters SITRANS MAG 8000 118 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 119: ... written printed or exported to a CSV file Only parameters data shown on white background can be changed Red texts are off line data which are not stored in the MAG 8000 yet whereas black texts show the actual meter data Each parameter has a description prompt advising how the parameter can be programmed and which settings are programmable The figure below shows alarm status with the marked alarms...

Страница 120: ...arious functional groups with maximum 99 parameters included For a complete parameter overview see Parameter lists in the appendix Page 136 Customized parameter list can be configured by creating a new parameter folder Flow Tool A 4 Setting the basic parameters SITRANS MAG 8000 120 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 121: ...r to the new folder These parameters are updated and handled as the original parameters and listed in the same order as copied to the customized parameter list Flow Tool A 4 Setting the basic parameters SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 121 ...

Страница 122: ...of parameters easier See also www siemens com flow www siemens com flow A 5 Unit selection MAG 8000 and MAG 8000 CT are delivered with totalizer and flow rate units as ordered via the MLFB structure The standard MAG 8000 units for each region are Europe m3 for totalizer and m3 h for flow rate US Gallon for totalizer and GPM Gallon per minute for flow rate Australia ML for totalizer and ML d Mega l...

Страница 123: ...s allowed A new sticker showing the new settings must then be affixed to the display Protected parameters The following parameters are protected see also parameter lists in appendix Page 136 MAG 8000 Standard MAG 8000 CT MAG 8000 Irrigation MAG 8000 CT additionally Parameter no Parameter name Parameter no Parameter name New password 101 Totalizer 1 5 Sensor tube diameter 102 Totalizer 2 7 Meter No...

Страница 124: ...key to PCB board for correct attachment see upper part of figure 5 Change units description in parameters 8 and 9 see Unit conversion table Page 135 6 Change units factor in parameters 300 and 301 see Unit conversion table Page 135 7 Change maximum flow rate Qn Q3 to new unit selection in parameter 10 8 Select display unit in parameter 306 9 Download each parameter to meter 10 Remove hardware key ...

Страница 125: ...t display information and accessible display menus Parameter 131 defines default display information with selection between Totalizer 1 Index 1 Totalizer 2 Index 2 Flow rate Index 3 updated with selected measuring frequency Fault codes Index 4 Customer totalizer Index 5 resettable Default information is shown after power up as well as after no key operation for 10 minutes Parameter 130 defines acc...

Страница 126: ...status Meter status parameter 120 gives a fast indication of reliability of revenue data It shows whether important information has been reset or manipulated for instance if meter has been powered down Status information can only be reset while hardware lock key is mounted Flow Tool A 8 Internal data handling SITRANS MAG 8000 126 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 127: ...to indicate alarms that have been active or that revenue data has been influenced in the specific period The logged information has a time and date stamp The data logger never stops storing data old data is overwritten following the first in first out principle i e log 1 is the last stored information which is moved to log 2 when the next logging is made and so on The consumption alarm indicates t...

Страница 128: ...ced with the real meter consumption every 4 hours The battery limit parameter 206 is the level at which the low power alarm is activated generating an alarm or call up if configured The power status parameter 513 follows the battery symbol on the display When switching between battery power from internal and external battery packs Battery power parameter 507 must be adjusted to match the actual nu...

Страница 129: ...curacy verification like the MAG VERIFICATOR FDK 083F5060 or FDK 083F5061 The MAG 8000 uses two main physical components for communication The IrDA sensor on the MAG 8000 An IrDA interface cable with USB or RS 232 connector B 1 Commissioning The qualification certificate is generated from PDM via the following steps 1 Enabling insulation test 2 Uploading the device data to the PC 3 Generating the ...

Страница 130: ...the Qualification Certificate The procedures for enabling the insulation test are shown below Set Insulation test enable to Yes if not already enabled Select Device Download to Device to download the change to the device Qualification certificate B 2 Enabling insulation test SITRANS MAG 8000 130 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 131: ...re available in section Insulation test in chapter Built in functions Page 87 B 3 Uploading the device data to the pc Select Device Upload to PC PG to load all device parameters from the device to the pc Qualification certificate B 3 Uploading the device data to the pc SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 131 ...

Страница 132: ... select Device Qualification Certificate The qualification certificate tool is protected by a service password that is only available for service specialists Qualification certificate B 4 Generating the qualification certificate SITRANS MAG 8000 132 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 133: ...k Generate qualification certificate in the next dialog box A progress bar shows the percentage of the process Qualification certificate B 4 Generating the qualification certificate SITRANS MAG 8000 Operating Instructions 11 2018 A5E03071515 AE 133 ...

Страница 134: ...status is of proper quality or not Reference values are available in the document Qualification Certificate Reference Guideline A5E02268573 See also Qualification Certificate Reference Guideline http support automation siemens com WW view en 66867834 Qualification certificate B 5 Result evaluation SITRANS MAG 8000 134 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 135: ... 0 001 BBL42 US oil barrel 6 289810743 Flow rate unit parameter 9 Correction factor parameter 301 Default 1 m3 s m3 min m3 minute 60 m3 h m3 hour 3600 m3 d m3 day 86400 GPS Gallon second 264 1721 GPM Gallon minute 15850 32 GPH Gallon hour 951019 4 GPD Gallon day 22824465 MGPD 1000000 Gallon day 22 824465 CFS ft3 second 35 31467 CFM ft3 minute 2118 882 CFH ft3 hour 127132 8 L s liter second 1000 L ...

Страница 136: ...nu L Statistic menu St Revenue menu R Factory settings marked in italic cannot be edited Note Change of units If the unit and factor is changed all unit related values become invalid It is recommended to reset all unit related values C 2 1 1 99 Param eter ID Meter version Display view Parameter data type Factory settings Data range Fixed parameter or meter data that are not changeable 1 All M1 App...

Страница 137: ...a that are not changeable 100 All M2 Actual date and time PS3 production date and time year month day T hours minutes seconds 101 All O1 Totalizer 1 0 0 to 2x109 102 All O2 Totalizer 2 0 0 to 2x109 103 All O5 Customer totalizer 3 0 0 to 2x109 104 All O5 Reset customer totalizer 3 No Yes No This parameter is removed from PDM parameter list The func tionality is implemented via PDM menu Device Total...

Страница 138: ...t log has been reset 7 Hardware key has been acti vated 8 Meter has been powered up If none of the above is active no flag is set 130 All Menu active 63 all menus active 0 to 63 binary presented with information 1 for bit 0 1 Operator menu 2 Meter info menu 3 Service menu 4 Log menu 5 Statistic menu 6 Revenue menu If none of the above is active the message is None active 131 All Default operator m...

Страница 139: ...message is No faults 201 All Alarm configuration list 254 Alarm 2 to 8 enabled 0 to 8191 See 200 If no alarm is enabled the mes sage is None enabled 202 All Date of fault log reset PS3 production date and time year month day T hours mi nutes seconds 203 All O4 Non optimal measure time in cluding the time of device run ning with insulation test fault coil current fault amplifier fault database chec...

Страница 140: ... T 00 00 00 220 All Amplifier alarm output enable Yes Yes No 221 All Amplifier fault hours 222 All Amplifier fault counter 223 All Amplifier fault appears 2000 01 01 T 00 00 00 224 All Amplifier fault disappears 2000 01 01 T 00 00 00 225 All Database alarm output enable Yes Yes No 226 All Database fault hours 0 227 All Database fault counter 0 228 All Database fault appears 2000 01 01 T 00 00 00 2...

Страница 141: ...fault disappears 2000 01 01 T 00 00 00 255 All Leakage alarm output enable No Yes No 256 All Leakage fault hours 0 257 All Leakage fault counter 0 258 All Leakage fault appears 2000 01 01 T 00 00 00 259 All Leakage fault disappears 2000 01 01 T 00 00 00 260 All Empty pipe alarm output ena ble No Yes No 261 All Empty pipe fault timer 0 262 All Empty pipe fault counter 0 263 All Empty pipe fault app...

Страница 142: ...de 1 15 Hz 1 60 Hz 1 30 Hz 1 15 Hz 1 5 Hz 1 5625 Hz 3 125 Hz 6 25 Hz 304 All Mains frequency MLFB dependent 50 or 60 Hz mains 305 All Decimal point MLFB dependent No point One digit after point Two digits after point Three digits after point Automatic point adjust 306 All Displayed unit Use m3 unit Displayed m3 unit or an arrow indica tion for a unit label 307 All Measurement direction MLFB depend...

Страница 143: ... 3 09 below MAG 8000 Irrigation 7ME6880 1 0 to 9 9 329 All Filter time constant 5 Tau 1 to 1000 331 All Excitation frequency limit 6 25 Hz for advanced version and 1 15 Hz for basic version 332 All Excitation frequency sensor limit Sensor related 7ME6810 and 7ME6820 6 25 Hz DN 25 to 200 1 to 8 3 125 Hz DN 250 to 600 10 to 24 1 5625 Hz DN 700 to 1200 28 to 48 7ME6880 3 125 Hz DN 25 to 600 1 to 24 1...

Страница 144: ...18 0 18 DN 80 3 0 17 0 68 0 68 0 68 0 27 0 27 0 27 0 43 0 17 0 17 DN 100 4 0 17 0 67 0 67 0 67 0 27 0 27 0 27 0 42 0 17 0 17 DN 125 5 0 17 0 66 0 66 0 66 0 27 0 27 0 27 0 41 0 17 0 17 DN 150 6 0 15 0 60 0 60 0 60 0 24 0 24 0 24 0 38 0 15 0 15 DN 200 5 0 17 0 68 0 68 0 68 0 27 0 27 0 27 0 42 0 17 0 17 DN 250 10 0 17 0 66 0 66 0 66 0 27 0 27 0 27 0 42 0 17 NA DN 300 12 0 24 0 61 0 61 0 38 0 24 0 24 ...

Страница 145: ...1 All M6 Baud rate 19 200 1200 2400 4800 9600 19 200 38 400 422 All M7 Parity Even 1 stop Even 1 stop Odd 1 stop None 1 stop None 2 stop 423 All Interframe space 35 35 to 255 424 All Response delay 5 1 to 50 ms 425 All Reset communication driver No Yes No This parameter is removed from PDM parameter list The func tionality is implemented via PDM menu Device Commu nication Settings Transfer The par...

Страница 146: ...ters stand by mode 3 As value 1 and 2 together 4 External power gone 5 As value 1 and 4 together 6 As value 2 and 4 together 7 As value 1 and 2 and 4 to gether 514 All Transmitter temperature Actual degree celsius 540 All Electrode impedance A Measured values 0 to 185 000 ohm 541 All Electrode impedance B Measured values 0 to 185 000 ohm 542 All Low medium impedance alarm 0 0 to 2 15x109 550 All C...

Страница 147: ...x109 to 1x109 603 All Limit for too low consumption 0 1x109 to 1x109 610 All L1 Date of latest log period 2000 01 01 T 00 00 00 year month day T hours mi nutes seconds 611 All L1 Latest Log period totalized 1 612 All Latest Log period totalized 2 0 613 All Latest Log period fault status No faults Active faults in log period 1 Insulation error 2 Coil current error 3 Amplifier overload 4 Database ch...

Страница 148: ...status See 613 619 All Log period 2 status informa tion See 614 620 All L3 Date of log period 3 621 All L3 Log period 3 totalized 1 622 All Log period 3 totalized 2 623 All Log period 3 fault status See 613 624 All Log period 3 status informa tion See 614 625 All L4 Date of log period 4 626 All L4 Log period 4 totalized 1 627 All Log period 4 totalized 2 628 All Log period 4 fault status See 613 6...

Страница 149: ...status See 613 654 All Log period 9 status informa tion See 614 655 All L10 Date of log period 10 656 All L10 Log period 10 totalized 1 657 All Log period 10 totalized 2 658 All Log period 10 fault status See 613 659 All Log period 10 status informa tion See 614 660 All L11 Date of log period 11 661 All L11 Log period 11 totalized 1 662 All Log period 11 totalized 2 663 All Log period 11 fault sta...

Страница 150: ... 16 fault status See 613 689 All Log period 16 status informa tion See 614 690 All L17 Date of log period 17 691 All L17 Log period 17 totalized 1 692 All Log period 17 totalized 2 693 All Log period 17 fault status See 613 694 All Log period 17 status informa tion See 614 695 All L18 Date of log period 18 696 All L18 Log period 18 totalized 1 697 All Log period 18 totalized 2 698 All Log period 1...

Страница 151: ...d 1 722 All Log period 23 totalized 2 723 All Log period 23 fault status See 613 724 All Log period 23 status informa tion See 614 725 All L24 Date of log period 24 726 All L24 Log period 24 totalized 1 727 All Log period 24 totalized 2 728 All Log period 24 fault status See 613 729 All Log period 24 status informa tion See 614 730 All L25 Date of log period 25 731 All L25 Log period 25 totalized ...

Страница 152: ...west fixed limit 811 Advanced Leakage source Flow rate Flow rate volume 812 Advanced Start period for leakage detection 120 min 2 00 24 00 0 to 1440 detection minutes 0 to 23 50 813 Advanced Duration leakage detec tion 0 0 to 144 10 to 1440 minutes 814 Advanced Leakage value unit Flow volume unit The unit is updated once param eter 811 change is downloaded to the device and afterwards up loaded to...

Страница 153: ...t to No when up loading device parameters to pc after command execution 821 Advanced St1 Latest leakage period flowrate 0 822 Advanced St1 Latest leakage period vol ume 0 823 Advanced Lowest measured leak age value 1 000 000 000 The unit is updated once param eter 811 change is downloaded to device and afterwards uploa ded to pc 824 Advanced Date of lowest leakage value 2000 01 01 T00 00 year mont...

Страница 154: ...dvanced R4 Tariff4 volume 4 0 853 Advanced R4 Tariff4 period time end 1080min 18 00 24 00 0 to 1439 minutes 23 59 854 Advanced R4 Tariff4 upper limit range 60 0 to 100 Qn Q3 855 Advanced R5 Tariff5 volume 5 0 856 Advanced R5 Tariff5 period time end 1260 min 21 00 24 00 0 to 1439 minutes 23 59 857 Advanced R5 Tariff5 upper limit range 80 0 to 100 Qn Q3 858 Advanced R6 Tariff6 volume 6 0 860 Advance...

Страница 155: ...te of consump tion profile PS3 production date and time year month day T hours mi nutes seconds 881 Advanced Reset consumption profile No Yes No Value is reset to No when up loading device parameters to pc after command execution 882 Advanced Total time in CP range 1 0 883 Advanced Upper limit in CP range 1 15 0 to 100 Qn Q3 884 Advanced Total time in CP range 2 0 885 Advanced Upper limit in CP ra...

Страница 156: ...200 1 to 48 The following table shows the relationship between flow velocity V flow quantity Q and sensor dimension DN ORZ YHORFLW Figure C 1 Sizing table Appendix C 3 Sizing sensor SITRANS MAG 8000 156 Operating Instructions 11 2018 A5E03071515 AE ...

Страница 157: ... I D inch or V Q x 283 67 Pipe I D 2 V ft s Q MGD Pipe I D inch C 4 Certificates You can find certificates on the Internet at AUTOHOTSPOT or on an included DVD See also Certificates on the internet http support automation siemens com WW view en 10806951 134200 C 5 Spare parts Accessories C 5 1 Ordering of spare parts Ensure that your ordering data is not outdated The latest ordering data is always...

Страница 158: ...sulation test Yes Leakage detection Yes Meter utilization Yes Statistics Yes Tariff Yes Settle date Revenue Yes Features Application identification parameters 1 and 2 Tag number visible on display if numbers are selected and meter location up to 15 characters per information Time and date parameter 100 Real time clock and date max 15 minutes drift per year Totalizer parameters 101 102 and 103 3 to...

Страница 159: ... DN 15 1 to 6 1 15 Hz 1 15 Hz DN 200 to DN 600 8 to 24 1 30 Hz 1 15 Hz DN 700 to DN 1200 28 to 48 1 60 Hz NA Excitation frequency with mains power follows maximum sensor excitation frequency Filter constant as numbers of excitations Low flow cut off of Qn Q3 Empty pipe detection active symbol on display when active Filter selection for mains power frequency 50 60 Hz Correction factor for change of...

Страница 160: ...3 Q4 Pulse overflow on output A and B Selected pulse volume is too small compared to actual flow rate and max output pulse rate Consumption Saved data logger consumption exceeds customer selected limit on high or low consumption Leakage Leakage detected based on customer settings Advanced version only Empty pipe No water in the pipe sensor Low impedance Measured electrode impedance below customer ...

Страница 161: ... chain for correct scaling Number of sensor measurements excitations Transmitter temperature battery capacity calculation Low impedance alarm for change in media Flow alarm when defined high flow exceeds Verification mode for fast measure performance check Advanced version includes Insulation Cross Talk test Meter utilization Consumption profile Statistical flow and consumption data Insulation tes...

Страница 162: ...Tariff values visible on the display Settling date Advanced version only At midnight 23 59 59 on a predefined date the totalizer 1 index value is stored Old values are stored to show the latest two totalized 1 index values Settling values are visible on the display Statistic Advanced version only Min flow rate with time and date registration Max flow rate with time and date registration Min daily ...

Страница 163: ...ote 17 Internet Contact person 8 83 Flow documentation 8 Support 82 Introduction 7 Items supplied 7 L Laws and directives 11 Lithium batteries Safety 12 M Mains supply 32 Maintenance 75 Material compatibility 11 P Parameter lists 136 PDM Commissioning 41 Pressure specifications 17 R Repair 75 Return procedures 83 S Safety 11 Instrument safety standards 11 Safety instructions Electrical connection ...

Страница 164: ...Index SITRANS MAG 8000 164 Operating Instructions 11 2018 A5E03071515 AE ...

Отзывы: