6.6
Data protection
Using hardware key
A hardware key is installed in the HL hole to change protected parameters. The
HL hole is located in the front of the PCB board behind the battery.
MAG 8000 Standard / MAG 8000 CT / MAG 8000 Irrigation)
Parameter no. Parameter name
-
New Password
5
Sensor tube diameter
7
Meter No.
8
Totalizer unit
9
Flow unit
10
Qn (Q3)
300
Totalizer unit factor
301
Flow unit factor
302
Pipe size
321
Calibration date
323
Calibration factor
325
Sensor offset
332
Max. sensor excitation frequency
Additional protected parameters (MAG 8000 CT)
Parameter no. Parameter name
101
Totalizer 1
102
Totalizer 2
303
Operation excitation frequency
305
Decimal point place
310
Flow direction totalizer 1
312
Flow direction totalizer 2
327
Adjustment factor
328
Low flow cut-off
332
Empty pipe impedance
420
Device Communication Address
550
Coil current active
551
Fix flow mode active
Commissioning
6.6 Data protection
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
55
Содержание SITRANS F M MAG 8000
Страница 10: ...Introduction 1 3 Further Information SITRANS MAG 8000 10 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 16: ...Description 3 4 Benefits SITRANS MAG 8000 16 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 56: ...Commissioning 6 6 Data protection SITRANS MAG 8000 56 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 74: ...Operation 7 6 Battery powered operation SITRANS MAG 8000 74 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 90: ...Troubleshooting FAQs 9 3 Flow simulation SITRANS MAG 8000 90 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 164: ...Index SITRANS MAG 8000 164 Operating Instructions 11 2018 A5E03071515 AE ...