
A.9
Battery configuration
Battery figures (generated as customer parameter list, see section Customer-selected
parameter list in chapter Setting the basic parameters show power management information.
Figure A-5
Battery configuration
At each battery replacement the capacity is reset to 100% (parameters 508 to 510) which is
then reduced with the real meter consumption every 4 hours.
The battery limit (parameter 206) is the level at which the low power alarm is activated
generating an alarm or call-up (if configured). The power status (parameter 513) follows the
battery symbol on the display.
When switching between battery power from internal and external battery packs, "Battery
power" (parameter 507) must be adjusted to match the actual number of batteries connected.
Flow Tool
A.9 Battery configuration
SITRANS MAG 8000
128
Operating Instructions, 11/2018, A5E03071515-AE
Содержание SITRANS F M MAG 8000
Страница 10: ...Introduction 1 3 Further Information SITRANS MAG 8000 10 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 16: ...Description 3 4 Benefits SITRANS MAG 8000 16 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 56: ...Commissioning 6 6 Data protection SITRANS MAG 8000 56 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 74: ...Operation 7 6 Battery powered operation SITRANS MAG 8000 74 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 90: ...Troubleshooting FAQs 9 3 Flow simulation SITRANS MAG 8000 90 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 164: ...Index SITRANS MAG 8000 164 Operating Instructions 11 2018 A5E03071515 AE ...