Data logger / Consumption alarm
The integrated data logger has 26 logging periods in which data can be stored daily, weekly
or monthly. The logger stores the consumption for totalizer 1 and totalizer 2 in the selected
period. Forward consumption is stored as a positive value and reverse consumption is stored
as a negative value. Alarm and meter status are also stored for the same period to indicate
alarms that have been active, or that revenue data has been influenced in the specific period.
The logged information has a time and date stamp. The data logger never stops storing data
- old data is overwritten following the first in/first out principle, i.e. log 1 is the last stored
information which is moved to log 2 when the next logging is made and so on.
The consumption alarm indicates that the actual consumption on totalizer 1 is above or below
the consumption limits.
Flow Tool
A.8 Internal data handling
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
127
Содержание SITRANS F M MAG 8000
Страница 10: ...Introduction 1 3 Further Information SITRANS MAG 8000 10 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 16: ...Description 3 4 Benefits SITRANS MAG 8000 16 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 56: ...Commissioning 6 6 Data protection SITRANS MAG 8000 56 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 74: ...Operation 7 6 Battery powered operation SITRANS MAG 8000 74 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 90: ...Troubleshooting FAQs 9 3 Flow simulation SITRANS MAG 8000 90 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 164: ...Index SITRANS MAG 8000 164 Operating Instructions 11 2018 A5E03071515 AE ...