Contents

0 - 3

Siemens AG

 

SPR2-230.061.01

Page  3 of  8

SIREMOBIL Iso-C

Medical Solutions

Rev. 13

12.04

CS PS 24

System Manual

 

_______Prerequisites __________________________________________________ 1 - 1

Required documents   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 1
Required tools, measurement and auxiliary devices   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 1
Information regarding this document  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 2

Position designations  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 2

Safety Information .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 3

Notes on the protective conductor resistance test .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 7
Notes on device leakage current measurement .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 1 - 8

Basic system   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 11

Removing the cover  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 11
Installing the cover.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 11

Cover over the horizontal carriage   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 12

Removing the cover  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 12
Attaching the cover   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 13
SIREPHOS cover  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 14

Monitor trolley  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 15

Removing the covers   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 15
Installing the covers  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 16

Installing/removing the cover at the C-arm mounting  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 16
Telescoping cover underneath the horizontal carriage   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 16
I.I. service position   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 17

Moving the I.I. into the service position  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 17
Installing the I.I.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 17

Concluding steps  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  1 - 18

_______Line voltage adjustment _________________________________________ 2 - 1

Checking the line voltage on the monitor trolley .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 2 - 1

_______Checking the operating voltages __________________________________ 3 - 1

Low voltages.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 3 - 1
I.I. Voltages  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .    . 3  -  1

_______Loading the operating software ___________________________________ 4 - 1

Control board D1, Host  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 4 - 1
Storage of the error log as ASCII file  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 4 - 1
Option 3D Reconstruction, Trace File and Event Log File.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 4 - 2

Saving trace files   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 4 - 2
Saving current trace files directly after display of an error .  .  .  .  .  .  .  .  .  .  .  .  .  . 4 - 3

_______Adjustments / Programs _________________________________________ 5 - 1

Checking the generator adjustment .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 5 - 3
Dose rate adjustment  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 5 - 5

Prerequisites.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 5 - 5

Page

Содержание SIREMOBIL Iso-C

Страница 1: ... 1998 The reproduction transmission or use of this document or its contents is not permitted without express written authority Offenders will be liable for damages All rights inclu ding rights created by patent grant or registration of a utility model _or_ design _are_ reserved SP System Manual Service Instructions ...

Страница 2: ...n levels Disclaimer The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services Assemblers and other persons who are not employed by or otherwise directly affiliated with or authori zed by Siemens or one of its af...

Страница 3: ...ng removing the cover at the C arm mounting 1 16 Telescoping cover underneath the horizontal carriage 1 16 I I service position 1 17 Moving the I I into the service position 1 17 Installing the I I 1 17 Concluding steps 1 18 2 _______Line voltage adjustment _________________________________________ 2 1 Checking the line voltage on the monitor trolley 2 1 3 _______Checking the operating voltages __...

Страница 4: ...at X Iris open 5 14 Full format X Iris closed 5 15 Zoom format X Iris closed 5 15 Zoom format X Iris open 5 15 Save values and end adjustment 5 16 Concluding tasks 5 16 Slot diaphragm 5 16 Disable circle mask 5 16 Start adjustment 5 16 Full format slot diaphragm open 5 16 Full format slot diaphragm closed 5 17 Zoom format slot diaphragm closed 5 17 Zoom format slot diaphragm opened 5 17 Enable cir...

Страница 5: ...30 6 2 Replacing the VIDEOMED DC 6 3 Camera rotation 0 position 6 3 Checks and adjustments 6 4 Basic system operating panel 6 4 Frequency converter FU Vertical lift 6 5 Power Supply M14 5 V 15 V 15 V 6 6 Power Supply M13 13 V VIDEOMED DC 6 6 I I Mini Voltage Supply 6 7 Collimator 6 10 SIREPHOS 6 11 Replacing the I I 6 12 Checking the temperature indicator 6 12 Removal 6 13 VIDEOMED DC 6 13 I I min...

Страница 6: ... reconstruction system electronics 6 39 Orbital motor 6 40 Orbital drive toothed belt 6 42 Orbital drive actual value potentiometer 6 44 Setting the actual value potentiometer for C arm orbital movement 6 45 Belt of the actual value potentiometer for orbital movement 6 46 Setting the actual value potentiometer for C arm orbital movement 6 46 Angulation part 6 47 7 ______ Brakes Movements _________...

Страница 7: ...age 7 of 8 SIREMOBIL Iso C Medical Solutions Rev 13 12 04 CS PS 24 System Manual 8 _______Supplement Measuring the tube current ___________________________ 8 1 9 _______Changes to previous version _____________________________________ 9 1 Page ...

Страница 8: ...0 8 Contents SIREMOBIL Iso C SPR2 230 061 01 Page 8 of 8 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions Page ...

Страница 9: ...g BENDER Unimet 1100 ST 51 38 727 Service PC with WINDOWS 3 1 3 11 or 95 operating system Service software refer to SIREMOBIL Iso C logbook Connection cable Service PC SIREMOBIL Iso C e g 99 00 440 1 set of resolution tests e g 28 71 820 1 set of radiation filters e g 97 98 596 Center cross e g 96 60 051 Heat conducting paste WPS e g 20 48 650 Special purpose oil Optimol Optipit 55 07 525 Special ...

Страница 10: ...ese service instructions will assist you in adjusting the SIREMOBIL Iso C for routine operation Not every adjustment is required each time the system is serviced Position designations 1 The designations used in these instructions left right front and back are explained in Fig 1 Fig 1 I I Mini power supply High voltage panel left right right front back back front left ...

Страница 11: ...ff and secure the power switch against accidental activation If an uninterruptible power supply UPS is installed in the system the power supply of the UPS must be switched free of voltage in addition or the voltage output plug must be pulled out If service in connection with electrical voltage is required the general safety information in instructions TD00 000 860 01 must be observed Electrical vo...

Страница 12: ...iance can lead to illnesses irreversible damage to body cells and the genotype severe injuries and even death In work on the system in which radiation must be released the radiation protection directives and the rules for radiation protec tion according to ARTD 002 731 02 must be complied with Note Use available radiation protection devices Wear radiation protection clothing lead apron Keep your d...

Страница 13: ...lat with fine emery cloth and close the damaged surface with paint A spray paint in the color of the C arm can be used for this purpose See Service Tool Catalogue Remove any carbon fibers lying in the C arm profile with a damp sponge or cloth If greater damage to the carbon fiber structure indicating persistent or large scale destruction of the carbon fiber structure e g crack formation or flaked ...

Страница 14: ...juries due to crushing cutting or grazing the skin especially to the hands Perform the corresponding work with special care and attentive ness Wear working gloves if necessary Danger of injury on mechanical parts Non compliance can lead to minor up to moderately severe inju ries especially to the hands With covers removed you can come into contact with parts such as for example flat plugs threaded...

Страница 15: ...measured documented and assessed The measurement must be made according to DIN VDE 0751 Part 1 see ARTD Part 2 In this case the protective conductor resistance to all conductive touchable parts of the system must be measured Make sure that control cables or data cables between the components of the system do not imitate any protective conductor connection During the measurement the mains cable and...

Страница 16: ...oltage line frequency before the measure ment Measurement Perform the measurement according to DIN VDE 0751 Part 1 see ARTD 002 731 17 and record the value determined as the first measured value Measurement of the device leakage current according to the residual current procedure measuring arrangement according to Fig 2 must be given preference since no danger for the person performing the measure...

Страница 17: ...uring instrument Fig 4 Measurement circuit for direct device leakage current measurement in compliance with DIN VDE 0751 1 2001 10 Fig C5 for Protection Class I 1 System 2 Applied part if present 3 Measuring arrangement integrated in measuring instrument Electric voltage Noncompliance can lead to severe injuries and even death In the direct measurement of the device leakage current measu ring arra...

Страница 18: ...t e g repairs to the power on circuit or replacement of the line filter the device leakage current measu rement must be repeated The same measuring conditions as in the first measurement apply Record and assess the highest value determined in the repeat measurement in the exi sting device leakage current record This value must not exceed the permissible leakage current values according to DIN VDE ...

Страница 19: ...ove the screws from the cover Pull the cover back Unscrew the protective conductor from the cover Pull the cover completely off and remove it Installing the cover 1 Reattach the protective conductor to the back cover Push the cover forward Replace the cover and tighten the screws Electrical voltage See Chapter 1 Safety Information Disconnect the line voltage plug prior to removing covers WARNING ...

Страница 20: ...needed disconnect the protective ground wire the cable of the keypad brake buttons and the cable of the emergency stop button Fig 5 Fig 6 Danger of injury when dismantling assembling mechanical parts Non compliance can lead to minor up to moderately severe injuries especially to the hands The cover over the horizontal carriage is bent U shaped and must be pulled on the side to the outside when it ...

Страница 21: ... Fig 3 Check the operation of the emergency stop button and of the brake buttons Fasten the cover with the 6 Allen screws 1 Fig 2 MODULARIS only Tightening force of the Docking Plate 18 Nm must be observed see service instructions for MODULARIS Danger of injury when dismantling assembling mechanical parts Non compliance can lead to minor up to moderately severe injuries especially to the hands Whe...

Страница 22: ...arm cover the SIREPHOS cover and the C arm Loosen the 3 cover screws of the SIREPHOS cover Lift off the cover On completing service place the SIREPHOS cover on the SIREPHOS Refasten the SIREPHOS cover with the 3 cover screws Re attach the C arm cover and seal it all around with sealing compound Use a paper towel to wipe off excess sealing compound Refasten the C arm cover with the 4 cover screws T...

Страница 23: ... cover Remove the protective conductor from the cover Fig 7 Risk of injuries on mechanical parts If not observed slight to medium injuries especially to the hands can occur Observe the notes concerning the C arm carbon fiber structure Risk of injury Noncompliance can lead to light to medium injuries and or property damage The C arm covers contain counterbalance weights and weigh approx 6 kg Hold t...

Страница 24: ...e screws Reattach the protective conductor for the upper back cover plate Install the cover and tighten the screws Installing removing the cover at the C arm mounting 1 Remove the cover screws and remove both covers from the side After completing the procedure reinstall both covers in the reverse order Telescoping cover underneath the horizontal carriage 1 Remove the four lower screws Slightly loo...

Страница 25: ...s be low Angulation Hold the I I firmly and carefully remove it from the C arm Be careful not to damage the cable Installing the I I 1 Move the C arm into the vertical position with the I I below the SIREPHOS above Holding the I I firmly carefully lift it up Be careful of the cable Holding the I I firmly move the C arm so that the Sirephos is above and the I I is below Angulation Carefully insert ...

Страница 26: ...R2 230 061 01 Page 18 of 18 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 1 18 Prerequisites Concluding steps 1 After completing all work perform the protective conductor test refer to ARTD 002 731 17 ...

Страница 27: ...mera video printer or video recorder must be disconnected from the line voltage supply Check the secondary voltages for line voltage transformer T1 according to table 1 Electrical voltage See Chapter 1 Safety Information Disconnect the SIREMOBIL Iso C line voltage plug from test point to test point Voltage Comment Transformer T1 31 Transformer T1 32 190 V to 205 V cannot be adjusted Transformer T1...

Страница 28: ...1 Page 2 of 2 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 2 2 Line voltage adjustment Option Optical isolation line voltage cable N Optical isolation line L voltage cable 230 V to 246 V cannot be adjusted Tab 1 ...

Страница 29: ...est point Voltage Potentiometer D1 X31 0V D1 X30 5 V 5 0 V to 5 2 V M14 5 V Adj D1 X34 0V D1 X32 15 V 14 9 V to 15 1 V M14 15 V Adj D1 X34 0V D1 X33 15 V 14 8 V to 15 2 V M14 15 V Adj D30 X1 8 0V D30 X1 9 24 V 24 0 V to 29 6 V not adjustable D30 X21 1 0V D30 X21 10 24 V 22 8 V to 29 5 V not adjustable D30 X20 1 0V D30 X20 3 27 V 27 3 V to 27 8 V D30 R115 M13 S 0V M13 S 13 V 13 1 V to 13 3 V M13 TR...

Страница 30: ...SIREMOBIL Iso C SPR2 230 061 01 Page 2 of 2 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 3 2 Checking the operating voltages This page intentionally left blank ...

Страница 31: ...am is already started Select the menu Diagnostics Select the menu Error log Click on the button Get from unit Wait until the error log is transferred to the service PC Click on the button Export to file Select your hard disk drive for storage of the file For the file name please enter the current date the part number and the serial number of the unit as described in the example Store the file on t...

Страница 32: ...elect the local service menu and enter the service key Activate the button Copy Trace to Drive A see Fig 1 All available trace files are copied onto the floppy disk as Zip file On occurrence of an error the trace is filed date and time as Zip file in the operating system on drive C trace trace_Vxxx_ zip Fig 1 In the copying process all already available older files are also copied until a floppy d...

Страница 33: ...led under the drive C Trace SPETrace_00 TRC The current error trace can be saved via local service Utilities source escape to OS Insert an empty formatted floppy disk in drive A of the Syngo PC In Syngo select the local service menu and enter the password Activate the Utilities button Under Source activate Escape to OS In the Parameters line enter copy C traces trc A and then activate the Go butto...

Страница 34: ...PR2 230 061 01 Page 4 of 6 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 4 4 Loading the operating software Saving the Event Log files Put in Servicekey and confirm with OK Select EventLog Fig 3 Fig 4 ...

Страница 35: ...ns Rev 13 12 04 CS PS 24 System Manual Select Time Range Minimum 1 month Notice following settings Domain Application Severity Error oder All All contain all informations Facility Service Source Name SP_ISO Message Limit ca 1000 Extract to File error Select Go The errors will be listet Close EventLog Fig 5 ...

Страница 36: ...ens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 4 6 Loading the operating software Select Utilites Insert a disk Select Escape to OS Parameters copy c syngo service extract error a Select Go Error File will be written to disk Fig 6 Fig 7 ...

Страница 37: ...Release the radiation release button as soon as the message Stop monitoring appears in the ACTIONS line Click on Update Stop Perform the generator adjustment filament circuit learning program The kV offset adjustment has to be followed by the generator adjustment filament circuit learning program Always perform the kV offset adjustment prior to beginning the generator adjustment X ray radiation Se...

Страница 38: ...pears in the STATUS line after the generator adjust ment has been successfully completed The Stop monitoring message appears in the ACTIONS line Activate the radiation release button Click on Update Stop The kV offset adjustment must be performed prior to starting the generator adjustment If an error occurs during the generator adjustment switch OFF the SIREMOBIL Iso C and perform the kV Offset an...

Страница 39: ...nnel 1 to D1 X46 ground at 0V Set the sensitivity to 1V Div Connect oscilloscope channel 2 to D1 X48 ground at 0 V Set the sensitivity to 2 5V Div Set the time to 25 ms Div Set the oscilloscope on storage mode Release an exposure with the hand switch and store the oscillogram with the exposure Evaluate the stored oscillograms The tube current must reach the tolerance range of 13 mA 1 5 mA in the t...

Страница 40: ...djustments Programs Error examples Fig 1 kV act 20 kV 1 0 V mA act 2 5 mA 1 0 V relevant time period Tube current exceeds the upper limit preset filament current too high 13 mA 6 ms 14 5 mA 11 5 mA 3 ms Tube current exceeds the lower limit preset filament current too low 13 mA 6 ms 14 5 mA 11 5 mA 3 ms Fig 1a Fig 1b ...

Страница 41: ... the I I input screen and attach The grid remains on the I I Correction factors 5 Grid correction factor 23cm 9 I I 1 5 27 cm 11 I I 1 5 Dose measurement chamber correction factor for Dali and Nomex 1 06 AUTO REGULATION indicates the current status of the automatic adjustment active not active manual done BRIGHTNESS DEVIATION shows the difference in the actual brightness value minus the nominal br...

Страница 42: ...ing dose rate values are acceptable if the dose rate level is set at LOW in zoom format When the dose rate level is set at LOW in the zoom format the minimum position value minimum opening of the TV iris may be the same value as in the survey format Opening of TV iris col limator in zoom format opening of TV iris collimator in survey format When the dose rate level is set at LOW in zoom format the...

Страница 43: ...sage done appears in the AUTO REGULATION line The BRIGHTNESS DEVIATION value is shown in the BRIGHTNESS DEVIATION line and should be 0 3 Switch off fluoro Setting the dose rate for zoom format 5 refer also to notice on previous page KV STOP remains selected Center the dose measurement chamber on the I I input screen and attach Select the I I zoom format refer to I I ZOOM STATUS The KV values remai...

Страница 44: ...input screen Set the kV mA curve HC1 3mA plateau curve If necessary temporarily program the HC1 curve During automatic mode kV STOP is off check the dose rate for dose rate levels LOW MID HIGH for the survey and zoom format Take into account the correction fac tors for the grid and the dose measurement chamber If necessary temporarily change the dose rate levels and kV mA curves Tolerance margin f...

Страница 45: ...Read and record the value displayed on the dose rate meter If the value measured exceeds 9 R min the maximum tube current must be reduced If the value measured is less than 9 R min the tube current does not need to be reduced Switch fluoroscopy off Setting the maximum skin dose rate is necessary only for coun tries where DHHS regulations apply Fig 2 If the maximum tube current was already reduced ...

Страница 46: ...ammed in Value Click on Set Value Select Put to unit Read the menu guide on the service PC Select Save to file The value to be programmed has been transferred to the SIREMOBIL Iso C and tempo rarily stored to the disk Checking 5 After programming check the maximum skin dose rate again tube current displayed at 110kV in mA 9 R min tube current to be programmed in mA measured dose rate value in R mi...

Страница 47: ... I I Select 2 II Full Format Position on the service PC and click on Execute Close the status window by clicking on the OK button For the query Do you want to reset the Unit click No Switch to zoom format During fluoroscopy adjust the X Iris by opening and closing the X Iris keys so that the collimator blades are still visible along the edge of the I I Select 3 II Zoom Format Position on the servi...

Страница 48: ...kup At Remarks for Parameter enter X Iris set Click on OK The new position values for the collimator will be stored on the disk Return to the main menu with the ESC key Collimator slot diaphragm 5 Prerequisites 5 Camera rotation has been adjusted correctly Image reversal functions are disabled The SLOT DIAPHRAGM TYPE OF DIAPHRAGM parameter has been programmed for DISPLAY YES Adjustment 5 Connect t...

Страница 49: ...to disappear at the edge of the im age Switch off fluoro In the window Available Functions select FULL FORMAT POSITION and click Exe cute Close the status window by clicking on the OK button Respond to the query Do you want to reset the unit by clicking No Select the zoom format Switch on fluoro Open the slot diaphragm to where the blades begin to disappear at the edge of the im age Switch off flu...

Страница 50: ...e of the SIREMOBIL Start the service program Select the Adjustment Calibrations menu Respond to the query with Yes Select Screen Display COLLIMATOR in the Function Groups combo box Start adjustment 5 In the Available Functions window select Init and click on Execute The X Iris is au tomatically opened to full format and the slot diaphragm is opened to maximum aperture Full format X Iris open 5 Zoo...

Страница 51: ... format is automatically selected Zoom format X Iris closed 5 Briefly release fluoro The X Iris is displayed on the monitor Use the keys for opening and closing the X Iris to set the diameter of the superimposed circle to the size of the X Iris In the Available Functions window select Close Pos II Zoom Format and click on Execute Close the status window by clicking on the OK button For the query D...

Страница 52: ...rations menu Answer the query with Yes Select SCREEN DISPLAY SLOT DIAPHRAGM in the Function Groups combo box Disable circle mask 5 Switch off zoom In the Available Functions window select DISABLE CIRCLE MASK and click on Ex ecute Close the status window by clicking on the OK button For the query Do you want to reset the Unit click on No Start adjustment 5 In the Available Functions window select I...

Страница 53: ...format is selected automatically Apply fluoro briefly Open and close the slot diaphragm display via the keys Align the lines shown on the monitor with the front edges of the slot diaphragm blades In the window Available Functions select Slot Diaphragm Close Pos I I Zoom For mat and click Execute Close the status window by clicking on the OK button Respond to the query Do you want to reset the unit...

Страница 54: ... display Slot diaphragm in the Combo box Click Get from Unit Select Rotation 0 degree position The parameter in the Value Actual field can be changed according to the minimum and maximum values Then click Set Value Click Put to Unit After the SIREMOBIL Iso C has booted compare the 0 position of the slot diaphragm blades to the display on the monitor If necessary repeat the programming Click Save t...

Страница 55: ... to adjusting camera rotation Camera rotation should be set in the 0 position Fig 3 Fig 4 and Fig 5 shows the 0 position With the right camera rotation key the arrow on the key is point ing clockwise the marker on the monitor will rotate clockwise The image contents will rotate counterclockwise WIth the left camera rotation key the arrow on the key is pointing counterclockwise the marker superimpo...

Страница 56: ... rotation to the 0 position The image left right up down corresponds to Fig 3 Stop fluoro in the window Available Functions select 3 0 Degree Position and click on Execute For the query Do you want to reset the Unit click No Start fluoro and rotate the camera to the 180 position with the counterclockwise cam era rotation key The image contents will rotate clockwise Stop fluoro in the window Availa...

Страница 57: ... SIREMOBIL Iso C Start the service program Select the Adjustment Parameter menu In the Combo box Parameter Groups select Diamentor Click on Get from Unit The Parameter 1 Diamentor Existing must be on YES Select parameter 2 Constant Qt Kg In Value Actual field enter the parameterTw Click on Set Value Click on Put to Unit Click on Save to File Fig 6 This adjustment is used to calibrate the area dose...

Страница 58: ...e Start Diamentor Test and click on Execute Close the status window by clicking on the Ok button Respond to the query Do you want to reset the unit by clicking No Press the 0 key reset key for fluoroscopic timer The tested Tw value is displayed on the monitor of the SIREMOBIL Iso C This value must match the programmed CONSTANT Qt Kg value Tolerance 5 By pressing the 0 key reset key for fluoroscopi...

Страница 59: ...ement device 5 Prerequisites 5 The calibration of the area dose product measurement device has already been per formed Preparations 5 Attach and center the center cross on the I I Fig 6 Select survey format Select KV mA curve plateau HC1 Attach and center the small 1cm3 dose measurement chamber of the dose measure ment device Dali Nomex onto the image intensifier Fig 7 Release fluoroscopy briefly ...

Страница 60: ...measured dose KE should be high enough that the uppermost measurement range of the dose measurement de vice Dali Nomex is utilized fully if necessary reset the area dose product display on the SIREMOBIL and the dose measurement device and repeat the measurement with a different fluoroscopic time Record the measured dose Record the area dose product displayed on the SIREMOBIL monitor FDPa Calculati...

Страница 61: ...deviation between the measured area dose product and the displayed area dose product The deviation must be 0 3 Deviation displayed area dose product FDPa measured area dose product FDPg measured area dose product FDPg For deviations 0 3 the area dose product measurement device must be replaced Fig 8 Fig 9 Fig 10 Meas chamber Sirephos I I Center cross Sirephos I I ...

Страница 62: ...n Not for service invoke the necessary test images on the service PC Video frequency The currently programmed frame rate is shown in the parentheses to the right After downloading the MEMOSKOP or after replacement the MEMOSKOP will start up at the default video frequency 50 Hz If the monitors are configured for a different video fre quency you must perform blind configuration for the cor rect vide...

Страница 63: ...g the image rotation key For SIREMOBIL Iso C always set to YES Product name Allows you to change the MEMOSKOP name displayed during system booting Enter SIREMOBIL Iso C Diagnostics Allows you to invoke test routines for repair by the manu facturer of the MEMOSKOP The MEMOSKOP can be reset via Load factory default as required Subsequently you have to program accord ingly The same applies if you rep...

Страница 64: ...e programmed brightness contrast surve LUT is selected automatically for the subtraction mode when equipped with a MEMOSKOP C Sub Default LUT 1 Default Edge After switch on the programmed edge enhancement is selected automatically For MEMOSKOP SUB set on 1 Software compatibility Not implemented Must be set to No Unit serial number Enter the serial number of the SIREMOBIL Iso C when it is equipped ...

Страница 65: ...Without installed MEMOSKOP keyboard Use the MEMOSKOP service program to modify the configuration With installed MEMOSKOP keyboard Invoke Technical Setup on MEMOSKOP press the CTRL T keys Press the 2 key and press return The video frequency menu is selected Depending on the required video frequency press the right cursor key Press the right cursor key Press the left cursor key Roadmap K factor phas...

Страница 66: ...installed MEMOSKOP keyboard On the MEMOSKOP reconnect the video cable from OUT 1 to OUT 2 Press CTRL T at the same time on the MEMOSKOP keyboard TECHNICAL SETUP is called up and is now visible on the monitor Call up the menu Number of monitors by pressing the 3 key and subsequently the RETURN key on the MEMOSKOP keyboard With the and cursor keys set the number to 1 monitor and press the RETURN key...

Страница 67: ... necessary Check the display functions and setting of the blades on the monitor Adjust if neces sary Ensure that the area dose product measurement device where present is functioning correctly Ensure that the laser targeting device where present is functioning properly Test the FL PFL DR and direct exposure functions of the SIREMOBIL Iso C Electrical voltage See Chapter 1 Safety Information Make s...

Страница 68: ... are positioned correctly Perform the kV offset adjustment and the generator adjustment filament learning Test the FL PFL DR and direct exposure functions if there is a cassette holder Interface board D30 6 Replace interface board D30 Check the 27 V 24 V 24 V voltage Adjust it if necessary Refer to chapter 3 of these instructions Checking the operating voltages Check the radiation release format s...

Страница 69: ...er to the actual value potentiometer of the Compact optics at points 2 and 3 Turn the Compact optics by the clamping ring slowly until the ohmmeter displays 5000 ohms 40 ohms 0 Position Do not adjust the camera rotation any further Install the new Videomed DC at the 0 position and tighten the clamp The center of the plug is in the 0 position exactly centered to the I I housing Refer to Fig 1 Recon...

Страница 70: ...ol must allow for unhindered rotation over the entire rotation range Check the camera rotation and if necessary readjust it The camera should be able to rotate 220 3 or 220 3 Check this using the display Check the dose rate and if necessary readjust it Perform the IQ Quick test First check the resolution If necessary adjust the optical sharpness at the Compact optics If the Modularis system is pre...

Страница 71: ...r Disconnect all the cables at the frequency converter after you have documented this Replace the frequency converter after removing the Allen screws Refer to arrows Fig 2 Reconnect all the cables Reconnect the power plug Secure the cables with cable ties Reattach the cables with the cable clamps Perform a function test of the column vertical lift Fig 2 Electrical voltage See Chapter 1 Safety Info...

Страница 72: ...erating voltages Power Supply M13 13 V VIDEOMED DC 6 Remove the Plexiglas cover over the power supply Remove the 4 Allen screws Replace the power supply Check the power supply voltage and if necessary adjust it Refer to chapter 3 of these instructions Checking the operating voltages Electrical voltage See Chapter 1 Safety Information Switch system power supply off and disconnect the power plug Ele...

Страница 73: ...pprox 6 kg Hold the covers firmly when installing or removing them Before the C arm cover will be loosened the C arm must be moved to its middle orbital position If the C arm covers are loosened the C arm may not be moved to its orbital end positions I I or SIREPHOS at the orbital end posi tion Rotate the camera into the 0 position prior to beginning work to avoid having to readjust the camera rot...

Страница 74: ... side refer to Fig 3 Move the I I into the service position Refer to chapter 1 I I service position Remove both attachment screws for the control section in the I I flange of the C arm ar row Fig 4 Carefully remove the control section from the C arm and disconnect all the plugs Disconnect the protective conductor Remove the VIDEOMED DC camera from the I I Refer to Replacing boards Replacing compon...

Страница 75: ...on from the I I flange of the C arm Fig 5 Remove the high voltage section from the C arm Disconnect all the plugs Dissipate any residual voltage in the cables by shorting them to ground Install the new high voltage section Connect the control section securing it with the screws only after you have adjusted it Refer to Adjusting the I I mini voltage supply Remount the VIDEOMED DC and tighten the cl...

Страница 76: ... of the collimator display on the monitor and if necessary adjust it Check the 0 position of the camera rotation Refer to the chapter Replacing the VIDEOMED DC 0 Position Check the camera rotation function Reinstall the I I on the C arm Refer to chapter 1 I I service position Collimator 6 Replace the collimator Ensure that the collimator is functioning properly and is set correctly If necessary ad...

Страница 77: ...sure that all cables are routed properly Reconnect the SIREPHOS plug and secure it Reconnect the protective conductor using the ground screw Do not install the rubber stop yet Reinstall the collimator and connect it Reinstall the dose measurement chamber laser targeting device and connect them Perform the kV offset adjustment Perform the generator adjustment filament circuit learning program Ensur...

Страница 78: ...ding to the Adjustment Instructions for the Laser Targeting Device RXR2 130 032 01 If the MODULARIS system is present perform the isocenter check Replacing the I I 6 Checking the temperature indicator 6 Prior to installing the I I check the temperature indicator If the inner square field of the indicator is white the I I did not exceed the temperature range Remove the temperature indicator If the ...

Страница 79: ... nut from the anode cable Remove the plug for the cassette contact Disconnect all the cables protective conductor cable to VIDEOMED DC etc from the I I At the same time remove the cable ties or the cable clamps Pull all the cables inside the C arm Risk of injuries on mechanical parts If not observed slight to medium injuries especially to the hands can occur Observe the notes concerning the C arm ...

Страница 80: ...emove the centering disk attached above the grid Remove the grid Desolder the wires for the cassette contact Installation 6 Compact optics 6 Reattach the Compact optics to the new I I When doing this thread the anode wire through the recess in the optics The camera optics must lie against the eccentric screws Reinstall the O ring and the cap nut for the anode cable The cable length should be 70 mm...

Страница 81: ... camera rotation point 2 and 3 At 0 there should be a resistance value of 5000 ohms 40 ohms Place the VIDEOMED DC in the 0 position on the compact optics and secure it with the clamp Refer also to chapter 6 Replacing the VIDEOMED DC Miscellaneous parts 6 Mount the grid to the new I I Attach the wires for the cassette contact Reroute and reconnect all the cables protective conductor cables etc Chec...

Страница 82: ... the I I on the C arm flange Refer to chapter 1 I I service position Check the collimator display on the monitor and if necessary adjust it Perform the I Q Quick test Check the setting of the laser light localizer if present If necessary adjust it If the MODULARIS system is present perform the isocenter check Execute the country specific acceptances 16 Partial acceptance test DHHS ...

Страница 83: ...e I I mini voltage supply to drop Risk of injuries on mechanical parts If not observed slight to medium injuries especially to the hands can occur Observe the notes concerning the C arm carbon fiber structure Risk of injury Noncompliance can lead to light to medium injuries and or property damage The C arm covers contain counterbalance weights and weigh approx 6 kg Hold the covers firmly when inst...

Страница 84: ...n of the I I mini voltage supply Remove both attach ment screws at the C arm flange Refer to the arrows Fig 6 Disconnect the anode cable from the high voltage section Remove the O ring and the cap nut from the cable Compact optics 6 Remove the attachment screws of the Compact optics If necessary remove the retain er clip for the plug cable leads as well Remove the Compact optics Fig 7 When removin...

Страница 85: ...ubber gasket between the Compact optics and the I I housing Install the new Compact optics Reattach the retainer clip for the plug cable routing Route the anode cable through the recess in the Compact optics The Compact optics must lie against the eccentric screws When installing the Compact optics make sure that no dust or dirt particles contaminate the I I output or the Compact optics NOTE ...

Страница 86: ...the 0 position on the Compact optics and tighten the clamp Refer also to Fig 6 Reinstall the connection cable refer to arrow Fig 6 and reattach the cable with a cable tie in the recess of the VIDEOMED DC board When doing this make sure that the cable tie and the cable do not extend beyond the edge of the board Reconnect plug X1 of the VIDEOMED DC cable spool Miscellaneous cables 6 Reconnect all ca...

Страница 87: ... cross on the monitor refer to Fig 9 Rotate the camera in the 220 position Mark the center again on the monitor refer to Fig 9 Rotate the camera in the in 220 position Mark the center again on the monitor refer to Fig 9 Place an 8 mm washer over the three markings All three markings must be located within the 8mm diameter of the washer An 8 mm washer has a somewhat larger inner diameter of 8 4 mm ...

Страница 88: ...h 220 and 220 markings Re fer also to Fig 9 Switch fluoro on briefly and evaluate the position on the monitor According to the ratio of the I I input screen diameter to the I I output screen diameter a shift of the camera optics by approximately 0 1 mm corresponds to a position shift of the washer displayed on the monitor of more than 1 7 mm 23 cm I I Secure the camera optics again Erase the three...

Страница 89: ...n the inner diameter and should not touch the edge of the washer If necessary repeat the adjustment until the three markings fall within the 8 mm diameter Secure the camera optics Retighten the eccentric screws of the camera optics Checks and adjustments 6 Check the camera rotation and if necessary adjust it Check the setting of the radiation iris and if necessary readjust it Check the display of ...

Страница 90: ...irst number describes the MEMOSKOP type 2 CE 3 CE 100 C C MOD 4 C SUB C SUB MOD Read and record the parameters for MEMOSKOP C E if you are able to select User setup After replacing the Memoskop con figure the MEMOSKOP C E according to the parameters recorded Number of Monitors Configure 1 for SIREMOBIL with 1 monitor Configure 2 for SIREMOBIL with 2 monitors Hardcopy Check whether 5000 is programm...

Страница 91: ...figure No Operating mode Para meters Change the parameters programmed for the organ program upon customer request VCR output Configure to Monitor A Hardcopy Release For SIREMOBILwith 1 monitor Configure to monitor A For SIREMOBIL with 2 monitors Configure to monitor B Text monitor For SIREMOBIL with 1 monitor Configure to monitor A For SIREMOBIL with 2 monitors Configure to both monitors Split Scr...

Страница 92: ...OP type 2 CE 3 CE 100 C C MOD 4 C SUB C SUB MOD The image storage MEMOSKOP C E 100 is a new image storage that works only with the correct number of images For this rea son if the image storage was set to its default values Menu Diagnostics Load Factory Defaults the parameter Maximum image numbers must be set to 100 images Read and record the parameters for MEMOSKOP C E 100 if you can select User ...

Страница 93: ...Default LUT Configure 2 Program a different LUT curve only upon cus tomers request Default Edge Configure 1 Program a different Edge enhancement value only upon customers request Software backward com patible Configure No Operating mode Para meters Change the programmed parameters only upon customer request VCR output Configure to monitor A Hardcopy Release For SIREMOBIL with 1 monitor configure t...

Страница 94: ...ens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 6 28 Replacing boards replacing components With present System MODULARIS the parameters for the Litho program have to be set again according the Service instructions LITHOSTAR MODULARIS NOTE ...

Страница 95: ...an be read out in the 2nd line of the User Setup The first number describes the MEMOSKOP type 2 CE 3 CE 100 C C MOD 4 C SUB C SUB MOD Read and record the parameters for MEMOSKOP if you can select User setup After replacing the MEMOSKOP configure the MEMOSKOP according to the parameters noted Number of Monitors Configure 1 for SIREMOBIL with 1 monitor Configure 2 for SIREMOBIL with 2 monitors Hardc...

Страница 96: ... 3D Reconstruction or the option DICOM Bridge is installed or a MOD drive is present con figure Yes Otherwise configure No Enable Bridge If the option 3D Reconstruction or the option DICOM Bridge is installed or a MOD drive is present con figure Yes Otherwise configure No Delay 3D PC DICOM Bridge Only available of if option 3D Reconstruction is installed Configure 120 Interface to 3D PC Only avail...

Страница 97: ... set again according the Service instructions LITHOSTAR MODULARIS Check the software status of the newly delivered MEMOSKOP If the software status of the MEMOSKOP differs from that of the system the system software status must be downloaded to the MEMOSKOP The MEMOSKOP software status can be read out in the 2nd line of the User Setup The first number describes the MEMOSKOP type 2 CE 3 CE 100 C C M...

Страница 98: ...te external disk If the option 3D Reconstruction or the option DICOM Bridge is installed or a MOD drive is present con figure Yes Otherwise configure No Enable Bridge If the option 3D Reconstruction or the option DICOM Bridge is installed or a MOD drive is present con figure Yes Otherwise configure No Delay 3D PC DICOM Bridge Only available of if option 3D Reconstruction is installed Configure 120...

Страница 99: ...to monitor A Hardcopy Release For SIREMOBIL with 1 monitor configure to monitor A For SIREMOBIL with 2 monitors Configure to monitor B Text monitor For SIREMOBIL with 1 monitor configure to monitor A For SIREMOBIL with 2 monitors configure to both monitors Split screen Configure horizontally Configure vertically only upon customer request Display of time Configure Yes Set Date Time Set the current...

Страница 100: ...HOS cover Replace the area dose product measurement device Perform the calibration of the area dose product measurement device Refer to chapter 5 Calibrating the Area Dose Product Measurement Test the accuracy of the area dose product measurement device Refer to chapter 5 Testing the accuracy of the area dose product measurement device Refit the SIREPHOS cover Seal the SIREPHOS cover using sealing...

Страница 101: ...ufficient cable length in the keyboard compartment Using cable ties tie the cable incl keyboard cable to the spring steel band again at a distance of approx 10 cm Perform a function check of the mouse the external mouse and the keyboard Reattach the paneling Replacing the external mouse 6 Switch the system off Unplug the external mouse at the rear and replace it with the new external mouse Perform...

Страница 102: ...itcher Push the new video switcher into the holding plate and reconnect all connectors Screw the holding plate back onto the monitor trolley Tighten the two clamping screws again Switch the system on After the system run up the SIMOMED HM Monitor reference monitor should display the MEMOSKOP reference image If the run up process of the 3D reconstruction PC or the syngo interface appears the green ...

Страница 103: ... Loosen the 4 fastening nuts of PC board D210 Unplug all connectors Replace board D210 and fasten the new PC board with the 4 nuts Reconnect all connectors Connect the monitor trolley to the power supply again Switch the monitor trolley on and wait for the run up Check the cut off delay After the power OFF button has been pressed the monitor trolley must switch off after approx 3 minutes Fig 11 El...

Страница 104: ...nect all connectors Perform the startup of the 3D reconstruction PC as described in the Startup Instructions Chapter 4 Then insert the backup restore disk in the logbook compartment of the SIREMOBIL Iso C into the disk drive of the PC Start the syngo service software Select the Backup Restore menu and restore the data Perform a function check of the PC Perform the IQ quick test Check the DICOM fun...

Страница 105: ... is attached in front of PC board D1 in the basic unit Fig 13 Switch the system off Unplug all connectors Replace the module Reconnect all connectors Fasten the shielded cables in the shield connectors again ensuring good ground con tact of the shielding If cable ties have been removed attach the cables with cable ties again Perform a download of the control software as described in the Startup In...

Страница 106: ...es incremental encoder Loosen the 2 fastening screws of the mounting plate incl motor 1 Fig 15 Loosen the 2 fastening screws at the two lateral brackets 1 Fig 16 Carefully remove the mounting plate incl motor In doing so push the toothed belt off the pulley ensuring that the actual value potentiometer and its pulley are not damaged or bent Detach the motor from the mounting plate Screw the new mot...

Страница 107: ...ate incl motor In doing so place the toothed belt over the pul ley again ensuring that the actual value potentiometer and its pulley are not damaged or bent Retighten the toothed belt of the orbital drive 1 Fig 14 When depressed by thumb pressure on the opposite side the belt should give approx 2 to 3 mm Perform a function check of the orbital drive To do this perform a 3D scan ...

Страница 108: ...ables incremental en coder Loosen the 2 fastening screws of the mounting plate incl motor 1 Fig 18 Loosen the 2 fastening screws at the two lateral brackets 1 Fig 19 Carefully remove the mounting plate incl motor In doing so push the toothed belt off the pulley ensuring that the actual value potentiometer and its pulley are not damaged or bent Unscrew the two lateral brackets from the coupling uni...

Страница 109: ...ainst the centering disk and screw it back on with the two set screws 1 Fig 20 Reinstall the mounting plate incl motor In doing so place the toothed belt on the pulley again ensuring that the actual value potentiometer and its pulley are not damaged or bent Reconnect the motor and incremental encoder Reattch the two lateral brackets 2 Fig 19 For guidance first insert the top two screws 1 Fig 19 bu...

Страница 110: ...iometer at the front of the cou pling unit 1 Fig 21 Take the belt off the pulleys Remove the actual value potentiometer Unscrew the Allen screw of the large pulley and remove the pulley from the actual value potentiometer 2 Fig 21 Unscrew the securing bracket from the old potentiometer and attach it to the new potenti ometer Attach the large pulley to the new actual value potentiometer Reattach th...

Страница 111: ...the actual value potentiometer counterclockwise to the end stop Fig 22 Then turn it clockwise two turns Retighten the Allen screw of the large pulley Perform the mechanical settings according to the Startup Instructions Chapter 6 Prior to setting the actual value potentiometer the C arm must be in the orbital 95 position The orbital limit switch on the tube side is actuated If this is not yet the ...

Страница 112: ...Retighten the two fastening screws Setting the actual value potentiometer for C arm orbital movement 6 The C arm is in the 95 position See note Loosen the Allen screw of the large pulley 2 Fig 23 Using a screwdriver turn the actual value potentiometer counterclockwise to the end stop Fig 24 Then turn it clockwise two turns Retighten the Allen screw of the large pulley Perform the mechanical settin...

Страница 113: ...arm to the 0 position Disconnect the cable of the actual value potentiometer Loosen the fastening screws of the angulation part Replace the angulation part In doing so ensure that when the C arm is perpendicular the markings black lines on the two toothed wheels coincide Reconnect the cable of the actual value potentiometer Perform the mechanical settings according to the Startup Instructions Chap...

Страница 114: ...SIREMOBIL Iso C SPR2 230 061 01 Page 48 of 48 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 6 48 Replacing boards replacing components This page intentionally left blank ...

Страница 115: ...riage consistently when the brakes are released is 80 N When the brakes are applied the force must be 180 N Measuring the horizontal swivel force 7 Swivel force when the brakes are released max 35 Nm Swivel force when brakes are applied 100 Nm Fig 1 Risk of injuries on mechanical parts If not observed slight to medium injuries especially to the hands can occur Observe the notes concerning the C ar...

Страница 116: ... are applied the force required for movement when the C arm is in the vertical position is 100 Nm Measuring the angular force required for movement 7 The attachment point for the measurement is the railing next to the image intensifier where the C arm has an orbital position of 0 and an angulation of 0 When the brakes are released the force required for movement is between 20 and 30 N When the bra...

Страница 117: ...e the three attachment screws and remove the brake magnet Remove the plug connection for the brake magnet and insert the new brake magnet Place the new brake magnet back on the shaft and attach it with the three screws Reattach the cable with cable ties Reattach the cover over the C arm cable guide Reinstall the cover Make sure the protective conductor is connected properly Perform a function test...

Страница 118: ...he swivel brake replacement kit 56 44 955 includes the replacement instructions SPR2 230 841 02 Please use this instructions for performing the work steps Risk of burn See Chapter 1 Safety Information When the swivel brake is released the brake magnet heats up Be sure to switch off the power supply and let the brake magnet cool off before touching it CAUTION ...

Страница 119: ...lace the brake Remove plug D30 X8 Eject the contacts from plug D30 X8 Route the wires of the new vertical lift brake to board D30 and insert the contacts in the plug housing Fig 4 Fig 5 Risk of injury Noncompliance can lead to medium to severe injuries and or property damage To replace the vertical lift brake the SIREMOBIL Iso C must be placed on its side This requires 6 people To prevent damage t...

Страница 120: ...le is seated correctly in the gears Attach the motor with the screws Install the vertical lift magnet on the vertical lift motor Return the SIREMOBIL Iso C to the upright position Lubricate the spindle and the lifting column mounting Check the vertical lift function Also check that the vertical lift motor switches off at the end positions of the vertical lifting column limit switches Fig 6 Risk of...

Страница 121: ...slowly and carefully move to the end position by briefly tapping the lifting column switch on the control console of the basic unit The lifting column motor must switch off prior to reaching the end position and the lifting column must come to a full stop Move the lifting column approximately 10 cm away from the end position Now move the lifting column at full speed in the direction of the end pos...

Страница 122: ...itch Emergency off switch pressed vertical lift must not be possible Emergency off switch released vertical lift must be possible Lubricating the lifting column 7 Lubricate the spindle of the lifting column and the mounting of the guide cylinder once a year Spindle 7 Lubricate the spindle with approximately 10 20 ml of Optimol Optipit special oil Guide cylinder 7 Grease the guide cylinder with app...

Страница 123: ...nterbalance weight Repeat the procedure at the other end of the C arm Fig 9 Fig 10 Risk of injuries on mechanical parts If not observed slight to medium injuries especially to the hands can occur Observe the notes concerning the C arm carbon fiber structure Risk of injury Noncompliance can lead to medium to severe injuries and or property damage The C arm covers contain counterbalance weights and ...

Страница 124: ...vements After the counterbalance weights have been reinstalled on both sides move the C arm orbitally from one end to the other several times In the case of SIREMOBIL Iso C 3D additionally perform a scan If the belt tension is still insufficient repeat the procedure and place the belt into the next tooth to shorten it further ...

Страница 125: ... out the kV value produced while the current is being measured Calculate the tube current as follows e g At 110 kV an overall current of 5 275 mA is measured When measuring the tube current the distribution current must be derived from the measured value The distribution current depends upon the kV and is calculated according to Ohm s law The distribution resistance is 400 MOhm tube current mA ove...

Страница 126: ...SIREMOBIL Iso C SPR2 230 061 01 Page 2 of 2 Siemens AG System Manual Rev 13 12 04 CS PS 24 Medical Solutions 8 2 Supplement Measuring the tube current This page intentionally left blank ...

Страница 127: ...anual Changes to previous version 9 Chap 0 Cover page revision level table of contents revised Chap 1 to Chap 9 Revision level revised Chap 1 Required tools measurement and auxiliary devices new revised Safety information added Chap 6 and 7 Safety information added Chap 9 Changes to previous version revised ...