SIREMOBIL Iso-C
SPR2-230.061.01
Page 32 of 48
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
6 - 32
Replacing boards / replacing components
Language
Set for language of the country.
Display of realtime clock
Configure Yes. Select No only upon customer request.
Display of rotated images
Configure Yes.
Monitor zoom
Configure No.
Monitor split
Configure Yes.
Monitor text
Configure Yes.
Maximum image number
Configure 900.
Disable doserate selection
in user setup
Set to "No" when equipped with an area dose product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" in User Setup are vis-
ible. Otherwise set to "Yes".
Unit serial number
Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Product name
Program "SIREMOBIL Iso-C".
Software backward compati-
ble
Configure No.
Default program change-
able
Default: Yes.
Depending on country specific regulations configure Yes or
No.
Create external disk
If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
Enable Bridge
If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
Delay 3D-PC/ DICOM
Bridge
Only available of if option "3D Reconstruction" is installed.
Configure 120.
Interface to 3D-PC
Only available if option "3D Reconstruction" is installed.
Configure yes.
Default LUT
Configure 2. Only upon customers request program a
different LUT curve.
Default Edge
Configure 1. Only upon customers request program a
different Edge enhancement curve.
Default LUT for Subtraction
Configure 1. Only upon customers request program a
different LUT curve.
Subtraction K-Factor,
phase A
Program "32". Program a different value only upon cus-
tomer request.
Subtraction K-Factor,
phase B
Program "MD 2". Program a different value only upon cus-
tomer request.
Subtraction Time of phase A Program "4,5".
Program a different value only upon customer request.
Roadmap K-Factor, phase A Program "32" .
Program a different value only upon customer request.