SIREMOBIL Iso-C
SPR2-230.061.01
Page 8 of 10
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
7 - 8
Brakes / Movements
Emergency off switch
7
•
Loosen the cover above the horizontal carriage.
•
Unsolder the emergency off switch.
•
Replace the emergency off switch.
•
Solder the wires on the emergency off switch.
•
Reinstall the cover on the horizontal carriage.
•
Check functioning of the emergency off switch.
- Emergency off switch pressed: vertical lift must not be possible.
- Emergency off switch released: vertical lift must be possible.
Lubricating the lifting column
7
Lubricate the spindle of the lifting column and the mounting of the guide cylinder once a
year.
Spindle
7
•
Lubricate the spindle with approximately 10 - 20 ml of "Optimol, Optipit" special oil.
Guide cylinder
7
•
Grease the guide cylinder with approximately 5 - 10 ml of special bearing grease
"Slick Pac PTFE" .
•
Then move the lifting column through the entire travel range twice to ensure that the
grease is evenly distributed.
Fig. 8
Spindle