
Replacing boards / replacing components
6 - 25
Siemens AG
SPR2-230.061.01
Page 25 of 48
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
•
Configure the following parameters in User setup for MEMOSKOP:
•
Perform the IQ quick test.
Maximum image number
n.a.
Disable doserate selection
in user setup
Program for "No" when equipped with an area dose product
measurement device. The menu "Storage of Dose on disk"
and the menu "Dose on Hardcopy output" in User Setup are
visible.
Otherwise set to "Yes".
Unit serial number
Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Product name
Program "SIREMOBIL Iso-C".
Default program change-
able
Default: Yes.
Depending on country specific regulations configure Yes or
No.
Software backward com-
patible
Configure "No".
Operating mode Para-
meters
Change the parameters programmed for the organ program
upon customer request.
VCR output
Configure to Monitor A.
Hardcopy Release
For SIREMOBILwith 1 monitor: Configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor
For SIREMOBIL with 1 monitor: Configure to monitor A.
For SIREMOBIL with 2 monitors:
Configure to both monitors.
Split Screen
Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time
Configure Yes.
Set Date Time
Set the current date and time.
Hospital Name
Program the hospital name.
Storage of Dose on disk
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise, configure No.
Dose on Hardcopy output
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
With present System MODULARIS the parameters for the Litho-
program have to be set again according the Service Instructions
LITHOSTAR MODULARIS.
NOTE