SIREMOBIL Iso-C
SPR2-230.061.01
Page 26 of 48
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
6 - 26
Replacing boards / replacing components
MEMOSKOP C-E 100
6
•
Replace the MEMOSKOP C-E 100.
The attachment screws are accessible on the underside of the storage device.
•
Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
•
Use the MEMOSKOP Service Software to import the language file.
•
Set the MEMOSKOP C-E 100 to the correct video frequency and configure it for the
number of existing monitors.
Refer to chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors ."
•
Configure the following parameters in Technical Setup for MEMOSKOP:
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP.
The MEMOSKOP software status can be read out in the 2nd line
of the User Setup.
The first number describes the MEMOSKOP type.
(2=CE; 3= CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
The image storage MEMOSKOP C-E 100 is a new image storage
that works only with the correct number of images. For this rea-
son, if the image storage was set to its default values
(Menu "Diagnostics", "Load Factory Defaults"), the parameter
"Maximum image numbers" must be set to 100 images.
Read and record the parameters for MEMOSKOP C-E 100 if you
can select User setup. After replacing the MEMOSKOP, configure
the MEMOSKOP C-E 100 according to the parameters noted.
Number of Monitors
Configure 1 for SIREMOBIL with 1 monitor.
Configure 2 for SIREMOBIL with 2 monitors.
Hardcopy
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Edge 1 coef set
Check whether 3 is programmed, If not, program this value.
Edge 2 coef set
Check whether 8 is programmed, If not, program this value.
Motion detection setup
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Date / Time format
Set the country-specific format.
Keyboard
Set to the keyboard layout being used.
Language
Set for language of the country.
Display of realtime clock
Configure Yes. Select No only upon customer request.
Display of rotated images
Configure Yes.
NOTE
NOTE