Brakes / Movements
7 - 5
Siemens AG
SPR2-230.061.01
Page 5 of 10
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Vertical lift brake
•
Remove the basic unit cover.
•
Carefully place the SIREMOBIL Iso-C on its side.
•
Remove plug D30.X8 and eject the contacts.
•
Remove the three attachment screws for the vertical lift brake and replace the brake.
•
Remove plug D30.X8.
•
Eject the contacts from plug D30.X8.
•
Route the wires of the new vertical lift brake to board D30 and insert the contacts in the
plug housing.
Fig. 4
Fig. 5
Risk of injury!
Noncompliance can lead to medium to severe injuries and/or
property damage.
To replace the vertical lift brake, the SIREMOBIL Iso-C must be
placed on its side. This requires 6 people. To prevent damage to
the SIREMOBIL Iso-C or the floor, we recommend using several
wool blankets or corrugated cardboard or other suitable floor
covering.
WARNING