SIREMOBIL Iso-C 3D
SPR2-230.814.01
Page 32 of 42
Siemens AG
Rev. 05
08.04
CS PS 24
Medical Solutions
6 - 32
3. Part of activities independent of serial number
Cable no. 11, item number n.a.,
6
is included in system cabling 71 39 855
•
Connect plug 3D-PC I/O card (see Fig. 76) to the "Relay card" connector at the PC and
secure it with screws.
•
Route the end VU1 to video switcher 1 (see Fig. 69), from the left side toward the top.
•
Attach cable ends 1-3 to the Remote connector (Fig. 77),
1 to 1, 2 to 2, 3 to 3.
•
Route the end VU2 to video switcher 2 (see Fig. 69), from the right side toward the top.
•
Attach cable ends 1-3 to the Remote connector (Fig. 77),
1 to 1, 2 to 2, 3 to 3.
•
Connect the plug to "Remote Status" on video switcher 2 (see Fig. 69).
•
Connect the "PC out" plug to D210 X3.
Printer is available
6
•
If a printer exists, disconnect the Remote plug at the printer and connect it with the
coupling of the Memoskop. Route the cable upward to the left.
•
Connect the ’Printer Remote’ plug to the ’Remote’ printer connection.
No printer available
6
•
If there is no printer, connect the ’Printer Remote’ plug with the Memoskop coupling. If
required, insulate the metal housing with tape.
•
Plug D100.X1001 is not connected. Insulate the contacts, roll up the cable and attach it
using cable ties.
Fig. 77 Cable no. 11
Fig. 78 Video switcher, remote connector
1
2
3
connector 3D PC I/O