Kapitelüberschrift hier eintragen
4 - 13
Siemens AG
SPR2-230.814.01
Page 13 of 14
SIREMOBIL Iso-C 3D
Medical Solutions
Rev. 05
08.04
CS PS 24
•
When drilling, ensure that the drill is placed vertically against the plate.
•
Due to the step length of the drill, a small 26 mm diameter bridge remains on the inner
edge of the hole. Remove this bridge with a half-round file. The diameter drilled should
be 28 mm across the entire plate thickness. Use the file to remove the burr as well.
•
Also deburr the two 3.5 mm holes. This can be done with a drill having a diameter of
approx. 8 mm to 10 mm. Place the drill manually against the 3.5 mm holes from the
inside and brake the burr.
•
Remove all drill shavings outside and inside with a vacuum cleaner.
•
Varnish the 28 mm hole with a varnish applicator and allow it to dry.
•
Position the connector with holding plate (1/Fig. 27) from the inside.
•
The connector is positioned centrally to the 28 mm hole, the lid opens downward.
•
Fasten the connector with holding plate using the screws and lock washers supplied
(1/Fig. 28).
Fig. 27
Fig. 28
1
1
4
3
2