Actions up to serial number 1408
2 - 11
Siemens AG
SPR2-230.814.01
Page 11 of 14
SIREMOBIL Iso-C 3D
Medical Solutions
Rev. 05
08.04
CS PS 24
Cable No. 3 (Fig. 25)
2
•
Remove the synthetic cover at the frequency converter and pull the green connector
(see Fig. 27).
•
Insert connector M16FUB into the location vacated by the connector at the frequency
converter.
•
Connect plug M16FUA with the pulled green connector.
•
Insert cables M16K1 3 and 4 into contacts 3 and 4 at the emergency stop relay.
•
Route the cable properly and reattach the synthetic cover.
Cable No. 4 (Fig. 26)
2
•
Insert connector D200.X222 into X222 on pc boad D200 after it has been installed.
•
Insert the other end of the cable into contact 1 and 2 of the emergency stop relay
M16.K1.
Fig. 25 Cable no. 3
Fig. 26 Cable no. 4
Fig. 27 Frequency converter
Connector