3. Part of activities independent of serial number
6 - 21
Siemens AG
SPR2-230.814.01
Page 21 of 42
SIREMOBIL Iso-C 3D
Medical Solutions
Rev. 05
08.04
CS PS 24
•
Remove the keyboard tray again.
•
Add mouse and keyboard (Fig. 51).
•
Attach the two cables to the cable holders. The cable from the mouse to the cable holder
(
Ê
/Fig. 51) should be approximately 350 mm long. No particular length is required for the
keyboard cable. Both cables have to be wrapped with tape (Scotch tape) at the entrance
to the cable clamp.
•
Route the connection cables along the spring steel cross bands and secure them with
cable ties at 100 mm intervals (Fig. 52).
•
Install keyboard tray.
•
Attach spring steel cross bands (Fig. 53).
•
You should be able to pull out the keyboard tray
- by pressing on both sides along the outside of the tray
- by pushing against the center of the keyboard tray.
•
Attach the protective conductor to the protective conductor bar.
Fig. 52 Keyboard tray with mouse and keyboard
Fig. 53 Cabling completed
Fig. 54 Spring steel cross bands