SIREMOBIL Iso-C 3D
SPR2-230.814.01
Page 10 of 14
Siemens AG
Rev. 05
08.04
CS PS 24
Medical Solutions
2 - 10
Actions up to serial number 1408
"Emergency stop circuit 71 39 988" package
2
Cable no. 1
2
•
Lift the flaps and pull off the lower part of the emergency stop switch from cable 1
(Fig. 23).
•
Install the upper part of the emergency stop in the housing of the orbital carriage.
•
Reinsert the lower part.
•
Connect the connector.
Cable no. 2
2
•
Unplug the connector D30.X1 on PC board D30.
•
Connect it with connector D30.X1A of cable 2 (Fig. 24).
•
Plug connector D30.X1B of cable 2 into plug-in location D30.X1 on PC board D30.
•
Connect plug M16.X20 to plug M16.X20 of the cable harness just installed.
•
At the main cable harness
Plug M3.X30.1 into plug contact 1.
Plug M3.X30.3 into plug contact 3.
Fig. 23 Cable number 1