
2 - 1
Siemens AG
SPR2-230.814.01
Page 1 of 14
SIREMOBIL Iso-C 3D
Medical Solutions
Rev. 05
08.04
CS PS 24
Actions up to serial number 1408
2
•
Proceed as follows after the SIREMOBIL Iso-C has been moved to the respective room
and the foot brake has been enabled.
Cover of the X-ray unit
2
•
Remove the cover.
•
Remove the protective conductor.
Cover of the horizontal carriage
2
If LITHOSTAR MODULARIS option is installed:
•
Remove the cover (= 6 Allen screws) and pull connector M3.X1.
•
Unscrew the protective conductor.
•
Remove the brake assembly.
•
The assembly previously removed has to be integrated into the new cover.
•
Deinstall the cover plate. It will later be replaced by a new one (Fig. 1).
After the installation of the SIREMOBIL Iso-C 3D upgrade, the
external part of the docking plate must be reinstalled. Perform the
installation and setting of the switches as described in the
LITHOSTAR MODULARIS Service Instructions.
Fig. 1
Cover plate
NOTE