
SIREMOBIL Iso-C 3D
SPR2-230.814.01
Page 6 of 14
Siemens AG
Rev. 05
08.04
CS PS 24
Medical Solutions
2 - 6
Actions up to serial number 1408
Upgrade cable set, item number 71 39 996
2
•
Move the single wire (Fig. 14) on the right side from the top to the bottom by pulling it
through the horizonal carriage and the lifting column.
•
Connect the single wire to the wire at the bundled cables (Fig. 13).
•
Lay the cable set out on the floor (Fig. 13).
•
Wrap the connections of the two wires using Scotch tape.
•
Cover the tape with talcum powder, or use silicone or a similar type of lubricant.
•
Pull the wire upward toward the top at the entrance to the horizontal carriage. You will
need one person pulling the wire upward and a second person feeding the wire from
below (Fig. 16).
•
If necessary, push the existing bundle of cables in the recess of the horizontal carriage to
the left using a large screwdriver.
•
In case of blockage, pull the cables back a little, turn them and continue pulling them
toward the top.
•
Pull the cables up to the mark.
•
Remove the two wires.
Fig. 13 Upgrade cable set
Fig. 14 Single wire
Fig. 15 Wire on top
Fig. 16 Wire below
right side