Operation
MOBILETT
SPR8-230.621.01.03.02
45
of 106
Preparation
◆
Press the switch halfway, that is to the first position.
– The preparation cycle for exposure is activated. The green "Ready" lamp goes
off and does not light up again until anode rotation has reached full speed (after
approx. 2.5 s). This is indicated by four short "beeps". The generator is now
ready for exposure release.
Exposure
◆
Press the switch to the second position.
– The amber exposure indicator lights during the exposure. The indication lasts
somewhat longer than the actual exposure, in order to make the operator
aware of very short exposures. A longer beep indicates the end of the expo-
sure.
If exposure is not released within 15 s the preparation will automatically be inter-
rupted and display message ERR 25 appears.
It is also possible to press the button directly to the second position (3.). The
exposure is then released immediately after the anode rotation has reached full
speed, i.e. after approx. 2.5 s.
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...