Safety
MOBILETT
SPR8-230.621.01.03.02
9
of 106
Checks, Tests
Before using the equipment for examination, the user must ascertain that all
safety-relevant devices function properly and that the MOBILETT is ready for
operation.
The radiation indicator may light during operation only while the radiography
switch is operated.
Special attention must be given to checking the displays and pilot lights for
proper operation (visual check).
Perform the checks and tests according to the daily check list.
Warning
The radiation indicator glows although the radiation release switch is not acti-
vated!
Undesired radiation exposure possible
◆
Switch off the MOBILETT immediately.
◆
Please notify the Siemens customer service.