Radiation Calculation
MOBILETT
SPR8-230.621.01.03.02
65
of 106
Example: Calculation radiation dose
Exposure with U = 70 kV, Q = 20 mAs and source to skin distance, r = 1.05 m:
From
Diagram 1,
read the corresponding value of Y for U= 70 kV:
Y = 37
µ
Gy x m2/mAs
The radiation dose D is calculated with
Equation (2)
:
D = 37 (
µ
Gy x m2/mAs) x 20 (mAs) / (1.05)2 = 671
µ
Gy
The radiation dose at a distance of 1.05 m with tube voltage 70 kV and 20 mAs is
D= 671
µ
Gy.