Operation
MOBILETT
SPR8-230.621.01.03.02
43
of 106
Setting the Exposure Parameters
The exposure parameters are set by using the kV and mAs buttons.
For more detailed information see chapter Technical Data.
The exposure data are listed in chapter Dosimetry
(→
Page 93)
Exposure Points
The concept behind the exposure point system is to easily change the density or
contrast of the film. Therefore, the kV and mAs may appear to change in odd
steps. This will, however, allow logical density and contrast adjustments.
Exposure points are set by Siemens customer service. Examples can be found in
chapter Dosimetry
(→
Page 93)
Selecting exposure parameters
◆
Press the kV or mAs buttons briefly.
– The exposure value increases/decreases by one step.
◆
Hold the button pressed down for a period of time.
– This results in a continuous increase/decrease of the particular exposure
parameter.
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...