Operation
Operator Manual
SPR8-230.621.01.03.02
52
of 106
Driving the unit
If you drive the unit, please observe the following:
❏
The power cable must not be connected to a wall outlet.
Driving with motor power, MOBILETT XP Hybrid only
The driving control has a function designed to prevent the MOBILETT from accel-
erating uncontrollably. This can occur in e.g. slopes or if the MOBILETT is forced
to go faster. The motor power will be shut-off until the MOBILETT slows down
again. This is to prevent possible accidents where the operator loses control of
the MOBILETT.
◆
Check that the power cable is not connected to a wall outlet.
◆
Turn the mode selection switch to position battery operation.
– Main switch (key): On
The maximum incline for the motordrive is 7
°
.
As long as one of the motor operation keys is pressed, the system blocks radia-
tion release.
Caution
Risk of injuring the operator!
The system drive may shut off suddenly, causing the unit to brake
◆
Be careful if you push the system manually while the motor is on.
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...