The Exposure Point System
MOBILETT
SPR8-230.621.01.03.02
67
of 106
The Exposure Point System
The exposure point system allows to easily change the density or contrast on the
film. Therefore, the kV and mAs may appear to change in odd steps. Step data is
shown in chapter Technical Data
(→
Page 77)
Density and contrast relations
For example, pressing the kV+ button once (using whole exposure points), or
pressing the mAs+ button twice, will increase film density by 25%. If double
density is desired, press the kV+ button three times, or the mAs+ button six
times. Observe that changing the kV affects both the density and the contrast.
If an exposure yields accurate density but should be higher in contrast, it can
easily be adjusted. Pressing the kV- button once (using whole exposure points)
and the mAs+ button twice will yield the specific density and the desired
contrast.
If the kV is set to step in half exposure points, it will correspond to the mAs steps
(e.g. press kV+ button six times for double density).
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...