Functional and Safety Checks
MOBILETT
SPR8-230.621.01.03.02
73
of 106
Annual Maintenance
Maintenance every 12 months is necessary to keep the equipment safe and oper-
ational.
◆
We recommend that the maintenance is performed by the Siemens customer
service.
Check of the system
Check of system parts
Interval
Operations to perform
12 months
Mechanical test
12 months
Functional check
12 months
Checking of image quality
12 months
Checking of kV/mAs accuracy
12 months
Electrical safety checks
12 months
USA only:
Checking of reproducibility
Interval
Material
Material No.:
12 months
Collimator lamp
8392016
24 months
Battery pack (left) BK1/BK2
08392024
24 months
Battery pack (right) BK3/BK4
08392032
48 months
Cable winder complete
06508746
12 months
Batteries remote control (Option)
1 x 9 V alkaline
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...