Operation
Operator Manual
SPR8-230.621.01.03.02
40
of 106
Adjusting the radiation field
◆
Switch on the collimator light.
◆
Set the appropriate film focus distance, using the built-in measuring tape.
◆
Turn the collimator until the light field is parallel with the sides of the cassette.
◆
Place the detachable protractor on the cassette.
◆
Read the inclination angle and adjust it.
◆
Collimate the radiation field for the desired image size.
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...