Operation
MOBILETT
SPR8-230.621.01.03.02
41
of 106
Collimating the radiation field using the diagram
A diagram is helpful when using a filter.
(1) Cassette size
(2) Scale index
◆
Follow the horizontal line from the y-axis, that belongs to the size of the used
cassette, until it crosses the leaning line indicating the chosen Source Image
Distance (SID).
◆
Follow the vertical line and read off the value on the x-axle indicating the scale
index.
◆
Set the adjustment knobs for the collimator according to the scale index value.
– This adjustment gives a field size which corresponds to the size of the
cassette. In those cases when the object to expose is smaller than the size of
the cassette reduce the field size.
Never adjust the collimator to a field greater than needed for the investigated
area.
MOB00199
(2)
(1)
Содержание MOBILETT XP
Страница 1: ...SP Instructions for use MOBILETT XP XP ECO XP Hybrid The New Dimension in Mobile X ray Imaging ...
Страница 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Страница 41: ...Operation MOBILETT SPR8 230 621 01 03 02 39 of 106 Lateral Exposure Angle Adjustment MOB00288 MOB00289 ...
Страница 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Страница 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Страница 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Страница 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Страница 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Страница 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Страница 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...