Functional and Safety Checks
Operator Manual
SPR8-230.621.01.03.02
72
of 106
Monthly checks
Please perform the following safety checks monthly:
Labels
◆
Check the legibility of the labels according to section Labels
(→
Page 99)
◆
If any labels must be replaced, please notify Siemens customer service.
DAP Measuring System (Option)
The test displays a value, default is 100.0
µ
Gym2. If the test value is not between
80 - 120
µ
Gym2, the system should be calibrated.
◆
Press the Test button if you want to perform the system test.
The calibration of the chamber must be checked in accordance to national regu-
lations or laws.