Radiation Calculation
Operator Manual
SPR8-230.621.01.03.02
62
of 106
Example: Calculation DAP
Exposure with U = 85 kV, Q = 10 mAs and scale index values for collimator adjust-
ment Sx = 10 und Sy = 10:
Scale index value for collimator adjustment
From
Diagram 1
, read the corresponding value of Y for U= 85 kV:
Y = 57
µ
Gy x m2/mAs
10 1214
16
8
6
4
10
20
30
40
50
60
70
60
70
80
90
U=85
Y= 57