Monitoring
84
Instructions for Use Carina SW 3.n
Setting alarm limits
1
Press »
Select Menu
« button.
2
Press »
Alarms
« button.
Carina displays the alarm limits set.
The current measured value is displayed between
the upper and lower alarm limits.
To set:
z
Press the appropriate button.
z
Select the value using the rotary knob and con-
firm.
= Upper alarm limit
= Lower alarm limit
Alarm
limit
Setting range
Value set on
delivery
P
aw
high
10 to 55 mbar
25 mbar
MV high
2 to 40 L/min
15 L/min
MV low
0.1 to 39 L/min
3 L/min
f high
10 to 50 bpm
20 bpm
T
apn
5 to 60 seconds
15 seconds
T
disconn
Mask: 0, 15, 30, 60, 90, 120 seconds
Tube: 0, 15, 30, 60 seconds
15 seconds
WARNING
Set the alarm limits according to the patient's
condition. Alarm conditions that are not
adapted to the patient may lead to no or false
alarms.
035
1
50
7
2
537
Содержание Drager medical Carina
Страница 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 13: ...Instructions for Use Carina SW 3 n 13 Intended Use Intended Use Intended Use 14 Sub acute care definition 14 ...
Страница 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...