Operation
74
Instructions for Use Carina SW 3.n
Volume guarantee
Can be switched on in ventilation mode SPN-PS.
Tidal volume guarantee
Inspiratory plateau pressure is automatically regu-
lated between P
max
and PEEP in such a way that
the set tidal volume VT
is applied.
Set volume guarantee with the parameters:
– Tidal volume »
VT
«
– Pressure limitation »
P
max
«
Switching on VG
z
In menu 3 press the »
VG
« button.
z
Select and confirm »
On
« with rotary knob.
z
Set P
max
.
z
Set VT.
To switch off:
z
Select »
Off
« and confirm.
WARNING
Set P
max
in such a way that critical airway
pressures are avoided.
WARNING
In VC-SIMV and SPN-PS VG mode it is not pos-
sible to guarantee volumes if there is a sub-
stantial leak (e.g. due to a leaky mask).