MAINTENANCE AND REPAIRS
8
105
8020065-ZLP4/2017-08-10|SICK
Subject to change without notice
O P E R A T I N G I N S T R U C T I O N S | Master Data Analyzer
8.2.1
Replacing the hand-held scanner (only in operating mode with bar code scanning)
The hand-held scanner is available under part number
6050652
.
Disconnecting the connecting cable
1. Remove the defective hand-held scanner from its holder.
2. Press down on the retaining lug of the sealing ring.
3. Remove the connecting cable from the faulty hand-held scanner.
Fig. 53: Replacing components – Hand-held scanner
Connecting a replacement device
1. Insert the connecting cable into the replacement device until you hear the click of the
interlocking device.
2. Place the hand-held scanner in the holder of the MDA.
8.2.2
Replacing the display
A replacement display is available under part number
1071521
.
Fig. 54: Replacing components – Display (removing the cables)
▸
Before replacing the display, you should make a note of how the cables are assigned to
the connections.
▸
Unused connections are covered with protective caps. Leave these protective caps in
place.