25
8020216/19YN/V3-0/2020-12| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K
Subject to change without notice
PRODUCT DESCRIPTION
3
3.3.3
DUSTHUNTER SP100 Ex-3K sender/receiver unit
The sender/receiver unit comprises two main subassemblies:
●
Electronics unit
The assembly contains the optical and electronic subassemblies for sending the laser
beam and receiving the scattered light.
●
Measuring probe
The measuring probe is available in different versions and nominal lengths as well as for
various gas temperature ranges and defines the device variant (
sender/receiver unit”, page 39
).
Fig. 8: DHSP-T2VxxEX-3K
sender/receiver unit
Electronic housing
Protective tube
Control line connection
Flange with tube
Probe head with receiver optics
Cleaning opening for sender optics
Duct wall
Sight glass
Non-return valve G1/4“
Measuring volume
NL
Clearance to inner duct wall
NOTE:
Clearance to inner duct wall
The clearance between internal duct wall and sender/receiver unit measuring volume
should be at least 100 mm. For the high-temperature version, the clearance should be
between 100 mm and 140 mm so that the measuring device does not protrude too far
into the duct and is not warmed up unnecessarily by the hot sample gas.