8020216/19YN/V3-0/2020-12| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K
Subject to change without notice
CONFIGURING
9
9.5
System configuration
9.5.1
Application parameters
Assign the sender/receiver unit to the measuring location
The sender/receiver unit can be assigned explicitly to the respective measuring location.
The following steps are then necessary:
1
Start the SOPAS ET program and connect to the measuring system (
2
Select device file “DH SP100” and move it to the “Project Tree” window.
3
Enter password level 1.
4
Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”.
Fig. 40: SOPAS ET menu: DH SP100/Maintenance/Maintenance
5
Select the “Configuration / Application Parameters” directory and enter the desired data
in the “Mounting location” entry field under “Device Identification”.
Fig. 41: SOPAS ET menu: DH SP100/Configuration/Application parameters (top window)
The respective device version connected is displayed automatically