TROUBLESHOOTING
13.6
Control unit troubleshooting measures
13.6.1
Setting the laser alignment
In case the laser is no longer correctly aligned when checking the laser alignment (
“Checking laser alignment”, page 102
), adjust the laser alignment anew.
To be done:
As required.
Ex-condition:
Check must be performed outside the Ex-area.
24 V DC power supply must be available outside the Ex-area (
“Power supply without control unit”, page 100
).
Work steps:
1
Disconnect the entire device (control unit and sender/receiver unit) from the power
supply.
2
Loosen the flange screws. Remove the sender/receiver unit from the duct. If necessary,
allow to cool down.
3
Switch off the purge gas supply.
4
Remove the purge gas hose.
5
Disconnect the potential equalization line.
6
Disconnect the connection plug, for this purpose release the plug protection guard (
“Connection line and plug protection guard.”, page 32
).
7
Move the sender/receiver unit out of the Ex-zone into a safe area.
8
Place the enclosure rear side of the sender/receiver unit on a flat surface.
9
Loosen both screws (
à
) (SW 7) of the hood (
).
10
Remove the hood.
11
Loosen the six protection tube screws (
â
). Turn the protection tube lightly
and pull it off upwards.
12
Visually inspect O-rings (
,
) at this opportunity and replace if there is any
visible wear or damage (
see “Replacing the protection tube O-ring”, page 103
).
13
Connect and switch the sender/receiver unit power supply on.
14
Tighten or loosen the laser alignment adjustment screws (
), and
check the alignment (
see “Checking laser alignment”, page 102
) until the alignment is
correct.
15
Disconnect power supply.
16
Refit the probe protection tube. Fasten with the six protection tube screws (torque
5 Nm).
17
Fit the cover and tighten the screws (2 Nm).
18
Bring the sender/receiver unit to the measuring location.
19
Connect the potential equalization line.
20
Fit the purge gas hose and switch on the purge gas supply.
21
Fit the sender/receiver unit on the duct flange (torque 20 Nm).
22
Insert the connector plug and secure with the plug protection guard (
line and plug protection guard.”, page 32
23
Put the complete device back into operation (switch on voltage and check measured and
status values).