O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K
Subject to change without notice
CONFIGURING
9
4
Switch to the “Configuration / Application selection” directory.
5
The basic type of the sender/receiver unit connected is displayed in the “Connected
variant” window (field “Application selection”). Click “Save selection” to assign to the
control unit
.
Fig. 44: SOPAS ET menu: MCU/Configuration/Application selection
9.5.3
Factory settings
Table 16: Factory settings
The steps required to modify these settings are described in the following Sections. For this
purpose, the devices must be connected in SOPAS ET (
), the Level
1 password set and the “Maintenance” mode set.
Parameter
Value
Function check
Every 8 h; output of control values (every 90 s)
on standard analog output
Analog output (AO)
[mA]
Live Zero (LZ)
4
Measuring range end value
(MBE)
20
Current during maintenance
0.5
Current by malfunction
21 (optional 1)
Response time
60 seconds for all measured variables
Measured variable
Output on AO
Value at LZ
Value at MBE
Dust concentration [mg/m³]
1
0 200
Scattered light intensity
2
Coefficients set (only for dust concentration)
0.00 / 1.00 / 0.00