TROUBLESHOOTING
Symptom
Probable Cause(s)
Countermeasure
See:
Nothing is displayed on
the display panel.
The power cable is
disconnected.
The main switch on the
distribution panel is off.
The power supply voltage is
wrong.
Check the power supply
and voltage and make the
connections correctly.
The display doesn't
change when a sample
(item to be weighed) is
placed on the pan.
The pan has been displaced.
Set the pan correctly on the
balance.
The display fluctuates
and
(the stability
mark) does not appear
readily.
The balance has been installed
in an unstable environment.
Eliminate the effects of
vibration and air movement.
Install the balance on a robust
platform.
The pan supporter caps (TX
series large pan models only)
have come off.
Fit the pan supporter caps (TX
series large pan models only).
The protective in-use cover is
touching the pan.
When using the protective in-
use cover, make sure that it is
fitted snugly against the surface
of the balance body.
-
The glass door of the windbreak
is open (TW/TX series small
pan models only).
Close all the glass doors before
reading the display.
-
The weighing result is
not accurate.
Span calibration has not been
performed.
Perform span calibration.
Is the display at zero before
weighing?
Press
to set the
display at zero before weighing.
The units that you
want to use are not
displayed.
The units that you want to use
have not been set.
Ser the units that you want to
use for
Menu operations are
not possible.
Menu operation is locked.
Release the menu lock.
The WindowsDirect
communication function
cannot be used.
What to Do If….
For details, see
"Troubleshooting the WindowsDirect Communication Function".