122
Connecting to a PC (RS-232C)
10
CONNECTION AND COMMUNICATION WITH PERIPHERAL DEVICES
Application weighing
Command
Function
R
Cancels application weighing mode setting
Piece counting
PCS
o
Sets the piece counting (PCS) mode
o
: 1 to 5 mode numbers
UW
o
= XX.XXX
Sets the unit weight
XX.XXX : Setting value
UW
o
Reads the unit weight*
UB
o
= XXX
Sets the reference number of pieces
XXX : Reference number of
pieces value
UB
o
Reads the reference number of pieces
RECAL
Recalculates the unit weight
Percentage weighing
G
Switches between percentage (%) and gram units
%1
Sets the percentage weighing mode
"%" can also be used.
Formulation
M
Sets the formulation mode
* With the TWC**3L and TXC**3L, gram units are displayed in 0.0002 g intervals.
When a value that the balance cannot display has been set with command UW
o
, however,
the value set with the command is used for calculations.
Other functions
Command
Function
Comparator
TRGT
Establishes the target mode
TARGET = XX.XX
Sets the target in the target mode
XX.XX: Setting value
LIMIT = XX.XX
Sets the target range in the target mode
CHKW
Establishes the checkweighing mode
OVR.RNG = XX.XX
Sets the checkweighing range upper limit
value in the checkweighing mode
XX.XX: Setting value
UND.RNG = XX.XX
Sets the checkweighing range lower limit
value in the checkweighing mode
HI.LIM = XX.XX
Sets the pass range upper limit value in the
checkweighing mode
LO.LIM = XX.XX
Sets the pass range lower limit value in the
checkweighing mode
GO
Reads the results
[Response command]
HL (above "too heavy" range)
HI (too heavy)
OK (appropriate weight, pass)
LO (too light)
LL (below "too light" range)
System-related commands
Command
Function
ID = XXXX
Sets the balance ID
XXXX: Setting value
ID
Reads the balance ID
STATE
Outputs the setting details
Содержание TW223L