If "ERR C" is displayed…
The calibration check has failed for one
of the reasons given below.
There is too large a discrepancy
between the zero point of the
balance and the sensitivity.
A container has been placed on the
pan.
The pan is not on the balance.
There is too large a discrepancy in
the value of the internal weight.
Press
and redo the operation from
the beginning. If even on doing this the
same display reappears, calibrate the
internal weight
(
.
2
Press
"END" will be displayed. The sensitivity drift
is adjusted and the balance will return to the
weighing mode.
Caution
Instructions
If calibration doesn't end
normally and the balance
stops, do not move it nor
leave it as it is.
Moving the balance in such a
condition may cause failure
because the internal weight is not
held correctly.
Before moving the balance, be
sure to turn the power on and start
it up correctly (so that the internal
weight is correctly held).