Installation
1
BEFORE WEIGHING
2
Enter the weight value.
If necessary, change the weight value to match
the weight that will be used for calibration. If
there is no need to change it, proceed to step
3
.
(If necessary enter the weight value.)
^
"Entering Numerical Values", page 49
For details on the weight values that can be
entered, see
.
With models that don’t feature the windbreak
3
Place the calibration weight on the
pan.
Wait until the flashing weight value display
changes to a flashing zero.
*
If "ERR C" is displayed…
Span calibration was not completed for
one of the following reasons.
There is too large a discrepancy
between the zero point of the
balance and the sensitivity.
A container has been placed on the
pan.
The pan is not on the balance.
The wrong weight has been placed
on the pan.
No operation has been performed
within 60 seconds of the flashing
weight value or zero display.
Press
(
with the TXB series)
and repeat the operation from the
beginning.
4
Take the weight off the pan.
"END" will be displayed and the balance will
return to the weighing mode.
!3!0!0.0!0!0
!2!9!9.9!9!8
/
/
//
/
/
/
/
/
/
/
/
/ /
/ /
/ /
/
/
/
//
//
/
//
//
//
/
/
/
/
/
/
/ /
/
!3!0!0.0!0!0
!2!9!9.9!9!8
/
/
//
/
/
/
/
/
/
/
/
/ /
/ /
/ /
/
/
/
//
//
/
//
//
//
/
/
/
/
/
/
/ /
/
! ! !0.0!0!0
/ / /
/
/
/
/
//
//
//
//
/
/
/
/ /
/ / /
/
! ! !0.0!0!0
/ / /
/
/
/
/
//
//
//
//
/
/
/
/ /
/ / /
/
!E!N!D
!E!N!D