102
1
Press
in the weighing mode.
This opens the main menu.
2
Select the target mode.
[TOOLS]
[TARGT]
What is the current situation?
What do you want to do?
Stability
Mark
Target Mode
To Set / Update
To Cancel
Lit
ON
Press
and go to step
3
.
Press
and go to step
4
.
Unlit
OFF
Press
and go to step
3
.
Go to step
4
.
! !T!A!R!G!T
Check the presence or absence of the
stability mark.
! !T!A!R!G!T
Check the presence or absence of the
stability mark.
9
COMPARATOR FUNCTION
Target Mode
The comparator function performs a comparison between the weight reading and a reference value or
target value and displays the status of this comparison.
Comparator Function
You can select and use either of these modes according to the environment of use and application.
Target mode
After setting a target value and a tolerance
range with respect to that target value,
excesses and deficits in relation to the target
value are indicated by
,
and
(the comparator symbols).
Checkweighing mode
After setting the threshold values at the
upper and lower limits of the pass range,
when a sample is weighed a pass or fail
determination is indicated by
,
and
(the comparator symbols). An
out of range determination is indicated by all
comparator symbols OFF.
*
Before setting the comparator function
It can be used in combination with the application function mode
.
If you are already using the application function mode, read
"When the application function mode is set…"
.
The comparator function settings are retained even if the power is switched off.